Calibration Procedures
189
9.3.4. PERFORMING A MANUAL CALIBRATION CHECK WITH THE
INTERNAL SPAN GAS GENERATOR
Turn ON the CAL_ON_NO2 feature
SAMPLE
NOX STB= XXX.X PPB NOX= XXXX
<TST TST> CAL CALZ
CALS
SETUP
SAMPLE
NOX STB= XXX.X PPB NOX= XXXX
<TST TST> CAL
CALZ
CALS
SETUP
Set the Display to show
the NOX STB test
function.
This function calculates
the stability of the NO/NO
x
measurement.
Toggle TST> button until ...
SAMPLE
RANGE=500.0 PPB NOX= XXXX
<TST
TST>
CAL CALZ CALS
SETUP
ZERO GAS enters the analyzer via the
External Scrubber.
The ZERO and/or SPAN
buttons will appear at various
points of this process.
It is not necessary to press
them.
Wait until NOX STB
falls below 0.5 PPB.
This may take several
minutes.
SPAN GAS enters the reaction cell from the
internal NO
2
permeation tube.
Wait until NOX STB
falls below 0.5 PPB.
This may take several
minutes.
Record NO
X
, NO or NO
2
zero point readings
DO NOT press the ENTR button
Record NO
X
, NO, NO
2
span point readings
DO NOT press the ENTR button.
Turn OFF the CAL_ON_NO2 feature
Note
If the ZERO or SPAN buttons are not displayed, the measurement made is out of
the allowable range for a reliable calibration. See Section 12 for troubleshooting
tips.
06858E DCN7057
Summary of Contents for T200
Page 2: ......
Page 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Page 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Page 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Page 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Page 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Page 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Page 393: ...06858E DCN7057 D 7 ...