Teledyne ML – T101 Operation Manual
Operating Instructions
87
4.6.3.1.
Analog Output Signal Type and Range Span Selection
To select an output signal type (DC Voltage or current) and level for one output channel,
activate the ANALOG I/O CONFIGURATION MENU (see Section 4.6.3) then press:
DIAG AIO
CONC_OUT_2 RANGE: 5V
SET>
EDIT
EXIT
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
DIAG AIO
AOUTS CALIBRATED:
NO
< SET
SET>
CAL
EXIT
FROM ANALOG I/O CONFIGURATION MENU
DIAG AIO OUTPUT RANGE:
5V
0.1V 1V 5V
10V
CURR
ENTR
EXIT
Toggle to set the
signal level and
type of the
selected channel
DIAG AIO
CONC_OUT_2:5V, CAL
< SET SET>
EDIT
EXIT
Press
SET>
to select the
analog output channel to be
configured. Press
EDIT
to
continue
Pressing
ENTR
records the new setting
and returns to the previous menu.
Pressing
EXIT
ignores the new setting and
returns to the previous menu.
DIAG AIO OUTPUT RANGE:
10V
0.1V 1V 5V
10V
CURR
ENTR EXIT
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...