Operating Instructions
71
NOTE
On analyzers with the multigas option activated (see Sections 4.5.1 and 6.8) the titles
displayed on the instrument’s front panel during the previous operation appear as:
LOW range appears as SO2 RANGE
high range appears as H2S RANGE
As the instrument switches from H
2
S mode to SO
2
mode and back, only the reporting
range and analog output associated with the gas currently being measured will be active.
The reporting range and analog output for the gas no being measured will continue to
report the last valid reading.
4.4.4.6.
Auto Range Mode (AUTO)
In AUTO range mode, the analyzer automatically switches the reporting range between
two user-defined ranges (low and high). The unit will switch from low range to high
range when the H
2
S concentration exceeds 98% of the low range span. The unit will
return from high range back to low range once both the H
2
S concentration falls below
75% of the low range span.
When set up to measure a single gas (H
2
S or SO
2
), in
AUTO
Range mode the instrument
reports the same data in the same range on both the
A1
and
A2
outputs and automatically
switches both outputs between ranges as described above.
To select auto range mode and set the upper span limits for the high and low ranges, press
the following menu button sequence.
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
SETUP X.X
RANGE MODE:
SNGL
SNGL IND
AUTO
ENTR EXIT
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
EXIT
x 2 returns
to the main
SAMPLE
display
SETUP X.X
RANGE MODE:
AUTO
SNGL
IND AUTO
ENTR
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET
UNIT
EXIT
SETUP X.X
HIGH RANGE:
500.0 Conc
0 0 5 0 0 .0
ENTR
EXIT
SETUP X.X
LOW RANGE:
500.0 Conc
0 0 5 0 0 .0
ENTR
EXIT
Toggle the numeral
buttons to set the
LOW
and
HIGH
range value.
ENTR accepts the
new setting, EXIT
ignores the new
setting.
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...