Troubleshooting & Service
Teledyne ML - T101 Operation Manual
230
3
40
40 mV
0.4
2
4
4
60
60 mV
0.6
3
6
5
80
80 mV
0.8
4
8
6
100
100 mV
1.0
5
10
9.5.9.3.
Status Outputs
The procedure below can be used to test the Status outputs.
1.
Connect a cable jumper between the “-“ pin and the “
” pin on the status
output connector.
21.
Connect a 1000 Ω resistor between the +5 V and the pin for the status
output that is being tested.
Table 9-8. Status Outputs Check Pin Out
PIN
(left to right)
STATUS
1
System Ok
2
Conc Valid
3
High Range
4
Zero Cal
5
Span Cal
6
Diag Mode
7
Spare
8
Spare
22.
Connect a voltmeter between the “-“ pin and the pin of the output being
tested (Table 11-8).
23.
Under the DIAG > SIGNAL I/O menu (Section 4.6.1), scroll through
the inputs and outputs until you get to the output in question. Alternately
turn on and off the output noting the voltage on the voltmeter, it should
vary between 0 volts for ON and 5 volts for OFF.
9.5.9.4.
Control Inputs
The control input bits can be tested by the following procedure:
1.
Connect a jumper from the +5 V pin on the STATUS connector to the +5 V
on the CONTROL IN connector.
24.
Connect a second jumper from the ‘-‘ pin on the STATUS connector to
the A pin on the CONTROL IN connector. The instrument should switch
from SAMPLE mode to ZERO CAL R mode.
25.
Connect a second jumper from the ‘-‘ pin on the STATUS connector to
the B pin on the CONTROL IN connector. The instrument should switch
from SAMPLE mode to SPAN CAL R mode.
In each case, the T101 should return to SAMPLE mode when the jumper is removed.
9.5.10.
CPU
There are two major types of CPU board failures, a complete failure and a failure
associated with the Disk-On-Module (DOM). If either of these failures occur, contact
the factory.
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...