Instrument Maintenance
Teledyne ML - T101 Operation Manual
208
Figure 8-3. H2S - SO2 Converter Assembly
•
Wrap the band heater around the new replacement cartridge and tighten
the screws using a high-temperature anti-seize agent such as copper
paste. Make sure to use proper alignment of the heater with respect to
the converter tubes.
•
Replace the converter assembly, route the cables through the holes in
the housing and reconnect them properly. Reconnect the grounding
clamp around the heater leads for safe operation.
•
Re-attach the tube fittings to the converter and replace the insulation
and cover.
•
Replace the instrument cover and power up the analyzer.