MANUAL CALIBRATION CHECKS
Informal calibration checks, which evaluate but do not alter the analyzer’s response
curve, are recommended as a regular maintenance item, and in order to monitor the
analyzer’s performance.
The following section describes the basic method for manually checking the calibration of
the Model T101 analyzer in H
2
S measurement mode. The same method may be used to
manually check the calibration of the T101 analyzers configured for SO
2
measurement by
substituting SO
2
span gas for the H
2
S span gas listed. See Section 6.8 for instructions for
performing calibration checks on analyzers configured for multigas measurement mode.
To carry out a calibration check rather than a full calibration, follow these steps.
STEP ONE
: Connect the sources of zero air and span gas as shown in Figure 6.1.
STEP TWO
: Perform the zero/span calibration check procedure:
SAMPLE
H2S STB=XXX.X PPB
H2S =XXX.X
< TST TST >
CAL
SETUP
ACTION:
Record theH
2
S
concentration
reading.
ACTION:
Supply span gas to the instrument
SAMPLE
H2S STB=XXX.X PPB
H2S =XXX.X
< TST TST >
CAL
SETUP
ACTION:
Record the H
2
S
concentration
reading.
Wait until
H2S
STB
is below
0.5 ppb. This
may take
several minutes.
The value of
H2S
STB
may jump
significantly.
Wait until it falls
below
0.5 ppb. This
may take several
minutes.
SAMPLE
H2S STB=XXX.X PPB
H2S
=XXX.X
< TST TST >
CAL
SETUP
The
SPAN
button appears during the transition from zero
to span. You may see both buttons.
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST >
CAL
SETUP
Scroll the display to the
H2S STB
test function.
ACTION:
Supply the instrument with zero gas.
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...