Teledyne ML – T101 Operation Manual
Operating Instructions
135
To configure the parameters for a specific data parameter, press:
FROM THE EDIT DATA PARAMETER MENU
(see previous section)
SETUP X.X
0) PARAM=S2SCN1, MODE=AVG
PREV NEXT
INS DEL EDIT EXIT
SETUP X.X
PARAMETERS:S2SCN1
SET> EDIT
EXIT
SETUP X.X
SAMPLE MODE:AVG
<SET SET>
EDIT
EXIT
SETUP X.X
PARAMETERS: S2SCN1
PREV NEXT
ENTR
EXIT
Cycle through list of available
Parameters.
ENTR
accepts the new
setting and returns to the
previous menu.
EXIT
ignores the new setting
and returns to the previous
SETUP X.X
SAMPLE MODE: INST
INST AVG MIN MAX
EXIT
Press the key for the desired mode
SETUP X.X
PRECISION: 1
<SET SET>
EDIT
EXIT
SETUP X.X
PRECISION: 1
1
EXIT
Set for 0-4
SETUP X.X
STORE NUM. SAMPLES: OFF
<SET
EDIT
EXIT
SETUP X.X
STORE NUM. SAMPLES: OFF
OFF
ENTR EXIT
Turn ON or OFF
<SET
Returns to
previous
Functions
4.8.2.5.
Sample Period and Report Period
The DAS defines two principal time periods by which sample readings are taken and
permanently recorded:
SAMPLE PERIOD
: Determines how often DAS temporarily records a sample reading
of the parameter in volatile memory. The
SAMPLE PERIOD
is set to one minute by
default and generally cannot be accessed from the standard DAS front panel menu, but is
available via the instrument’s communication ports by using APICOM or the analyzer’s
standard serial data protocol.
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...