4.6.6.
LAMP CALIBRATION
An important factor in accurately determining H
2
S concentration, once the H
2
S is
converted to SO
2,
is the amount of UV light available to transform the SO
2
into SO
2
*
(see Sections 10.1.1 and 10.1.2). The Model T101 compensates for variations in the
intensity of the available UV light by adjusting the H
2
S concentration calculation using a
ratio (
LAMP RATIO
) that results from dividing the current UV lamp (
UV LAMP)
intensity by a value stored in the CPU’s memory (
LAMP_CAL
). Both LAMP Ratio and
UV Lamp are test functions viewable from the instrument’s front panel.
To cause the analyzer to measure and record a value for
LAMP_CAL
, press:
Exit at
any time
to return
to main
the
SETUP
menu
DIAG
SIGNAL I / O
NEXT
ENTR
EXIT
Repeat Pressing
NEXT
until . .
.
DIAG
LAMP CALIBRATION
PREV NEXT
ENTR
EXIT
DIAG FCAL
LAMP CAL VALUE:4262.4 mV
4 2 6 2 .4 ENTR EXIT
E
NTR
accepts the
new value
EXIT
ignores the new
value
The value displayed is the
current output of the UV
source reference detector
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > CAL
SETUP
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...