Teledyne ML - T100 UV Fluorescence SO2 Analyzer
Getting Started
75
3.7.1.4.
ZERO/SPAN CALIBRATION
To perform the zero/span calibration procedure, press:
Press ENTR to changes
the OFFSET & SLOPE
values for the SO
2
measurements.
Press EXIT to leave the
calibration unchanged and
return to the previous
menu.
SAMPLE
STABIL= XXXX PPB
SO2=XXX.X
< TST TST > CAL SETUP
M-P CAL
STABIL= XXXX PPB
SO2=XXX.X
<TST TST>
ZERO
CONC
EXIT
EXIT at this point
returns to the
SAMPLE menu.
Press ENTR to changes
the OFFSET & SLOPE
values for the SO
2
measurements.
Press EXIT to leave the
calibration unchanged and
return to the previous
menu.
SAMPLE RANGE=500.0 PPB SO2= XXXX
< TST
TST >
CAL SETUP
Set the Display to show
the STABIL test function.
This function calculates
the stability of the SO
2
measurement.
Toggle TST> button until ...
Allow zero gas to enter the sample port
at the rear of the analyzer.
Wait until STABIL falls
below 0.5 ppb.
This may take several
minutes.
SAMPLE
STABIL= XXXX PPB
SO2=XXX.X
< TST TST >
CAL
SETUP
M-P CAL
STABIL= XXXX PPB
SO2=XXX.X
<TST TST>
ENTR
CONC
EXIT
M-P CAL
STABIL= XXXX PPB
SO2=XXX.X
<TST TST> ZERO
SPAN
CONC
EXIT
Allow span gas to enter the sample port
at the rear of the analyzer.
Wait until STABIL falls
below 0.5 ppb.
This may take several
minutes.
SAMPLE
STABIL= XXXX PPB
SO2=XXX.X
< TST TST >
CAL
SETUP
M-P CAL
STABIL= XXXX PPB
SO2=XXX.X
<TST TST>
ENTR
CONC
EXIT
M-P CAL
STABIL= XXXX PPB
SO2=XXX.X
<TST TST> ENTR
CONC
EXIT
The SPAN key now appears
during the transition from
zero to span.
You may see both buttons.
If either the ZERO or SPAN
buttons fail to appear see the
Troubleshooting section for
tips.
Figure 3-28:
Zero/Span Calibration Procedure
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...