3.7. INITIAL CALIBRATION OF THE T100
To perform the following calibration you must have sources for zero air and span gas
available for input into the sample port on the back of the analyzer. Refer to Section 3.5
for instructions for connecting these gas sources.
The initial calibration should be carried out using the same reporting range set up as
used during the analyzer’s factory calibration. This will allow you to compare your
calibration results to the factory calibration as listed on the
Final Test and Validation
Data Sheet
, P/N 04551.
If both available DAS parameters for a specific gas type are being reported via the
instruments analog outputs (e.g.
CONC1
and
CONC2
when the
DUAL
range mode is
activated), separate calibrations should be carried out for each parameter.
Use the
LOW
button when calibrating for
CONC1
(equivalent to
RANGE1
).
Use the
HIGH
button when calibrating for
CONC2
(equivalent to
RANGE2
).
Refer to the Configurable Analog Output Addendum, P/N 06270 for more information
on the configurable analog output reporting ranges.
IMPORTANT
The following procedure assumes that the instrument does not have any of the available
Valve Options installed. Refer to Section 6.4 for instructions for calibrating instruments
possessing valve options
Note
The T100 analyzer has been tested for its ability to reject interference for most sources.
See Section 9.1.9 for more information on this topic.
3.7.1. INITIAL CALIBRATION PROCEDURE FOR BASIC ANALYZERS
(NO OPTIONS
The following procedure assumes that:
The instrument DOES NOT have any of the available calibration valve or gas inlet options
installed;
Cal gas will be supplied through the SAMPLE gas inlet on the back of the analyzer (refer to
Figure 3-4), and;
The pneumatic setup matches that described in Section 3.5.
3.7.1.1.
VERIFYING THE REPORTING RANGE SETTINGS
While it is possible to perform the following procedure with any range setting we
recommend that you perform this initial checkout using following reporting range
settings:
Unit of Measure: PPB
Analog Output Reporting Range: 500.0 ppb
Mode Setting: SNGL
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...