Specifications, Approvals & Compliance
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
30
Parameter
Description
Environmental
Installation category (over-voltage category) II; Pollution degree 2
Operating Temperature
Range
5 - 40
o
C (with EPA Equivalency)
Humidity Range
0 - 95% RH, non-condensing
Dimensions HxWxD
7" x 17" x 23.5" (178 mm x 432 mm x 597 mm)
Weight
31 lbs (14 kg)
35.7 lbs (16 kg) with internal pump
Certifications
EN61326 (1997 w/A1: 98) Class A, FCC Part 15 Subpart B Section 15.107 Class A,
ICES-003 Class A (ANSI C63.4 1992) & AS/NZS 3548 (w/A1 & A2; 97) Class A.
IEC 61010-1:90 + A1:92 + A2:95,
Approvals
US EPA: EQSA-0495-100
MCERTS: Sira MC 050067/04
EN14212, Approved, CE and Others
1
As defined by the USEPA.
2
Defined as twice the zero noise level by the USEPA.
Table 2-2: O
2
Sensor Option Specifications
Parameter
Description
Ranges
0-1% to 0-100% user selectable. Dual ranges and auto-ranging supported.
Zero Noise
1
<0.02% O
2
Lower Detectable Limit
2
<0.04% O
2
Zero Drift (24 hours)
3
<± 0.02% O
2
Zero Drift (7 days)
<±- 0.05% O
2
Span Noise
1
<
±
0.05% O
2
Span Drift (7 days)
<± 0.1% O
2
Accuracy
(intrinsic error) <± 0.1% O
2
Linearity
<± 0.1 % O
2
Temp Coefficient
<± 0.05% O
2
/°C,
Rise and Fall Time
<60 seconds to 95%
1
As defined by the USEPA
2
Defined as twice the zero noise level by the USEPA
3
Note: zero drift is typically <± 0.1% O
2
during the first 24 hrs of operation
Parameter
Description
Ranges
0-1% to 0-20% user selectable. Dual ranges and auto-ranging supported.
Zero Noise
1
<0.02% CO
2
Lower Detectable Limit
2
<0.04% CO
2
Zero Drift (24 hours)
<± 0.02% CO
2
Zero Drift (7 days)
<± 0.05% CO
2
Span Noise
1
<± 0.1% CO
2
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...