Troubleshooting & Repair
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
270
06807A(DCNxxxx)
Check for disconnected cables to the sensor module.
Carry out an electrical test with the
ELECTRICAL TEST
procedure in the diagnostics
menu, refer to Section 4.8.6. If this test produces a concentration reading, the analyzer’s
electronic signal path is working.
Carry out an optical test using the
OPTIC TEST
procedure in the diagnostics menu, refer to
Section 4.8.5. If this test results in a concentration signal, then the PMT sensor and the
electronic signal path are operating properly. If the M passes both ETEST and OTEST, the
instrument is capable of detecting light and processing the signal to produce a reading.
Therefore, the problem must be in the pneumatics, optics or the UV lamp/lamp driver.
10.4.3. UNSTABLE ZERO AND SPAN
Leaks in the T100 or in the external gas supply and vacuum systems are the most
common source of unstable and non-repeatable concentration readings.
Check for leaks in the pneumatic systems as described in Section 8.3.6. Consider pneumatic
components in the gas delivery system outside the T100 such as a change in zero air source
(ambient air leaking into zero air line or a worn-out zero air scrubber) or a change in the span
gas concentration due to zero air or ambient air leaking into the span gas line.
Once the instrument passes a leak check, perform a flow check (refer to Section 8.3.7) to
ensure that the instrument is supplied with adequate sample gas.
Confirm the UV lamp, sample pressure and sample temperature readings are correct and
steady.
Verify that the sample filter element is clean and does not need to be replaced.
10.4.4. INABILITY TO SPAN - NO SPAN BUTTON
In general, the T100 will not display certain control buttons whenever the actual value of
a parameter is outside of the expected range for that parameter. If the calibration menu
does not show a SPAN button when carrying out a span calibration, the actual
concentration must be outside of the range of the expected span gas concentration, which
can have several reasons.
Verify that the expected concentration is set properly to the actual span gas concentration in
the
CONC
sub-menu.
Confirm that the SO
2
span gas source is accurate.
If you are using bottle calibration gas and have recently changed bottles, bottle to bottle
variation may be the cause.
Check for leaks in the pneumatic systems as described in Section 10.6.1. Leaks can dilute the
span gas and, hence, the concentration that the analyzer measures may fall short of the
expected concentration defined in the
CONC
sub-menu.
If the physical, low-level calibration has drifted (changed PMT response) or was accidentally
altered by the user, a low-level calibration may be necessary to get the analyzer back into its
proper range of expected values. One possible indicator of this scenario is a slope or offset
value that is outside of its allowed range (0.7-1.3 for slope, -20 to 150 for offsets). Refer to
Section 10.7.2.8 on how to carry out a low-level hardware calibration.
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...