Theory Of Operation
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
238
06807A(DCNxxxx)
9.4.4. PNEUMATIC SENSORS
The T100 uses two pneumatic sensors to verify gas streams. These sensors are located
on a printed circuit assembly, called the pneumatic pressure/flow sensor board. The flow
simultaneously enters the sample pressure sensor and the flow sensor from the outlet of
the reaction cell.
9.4.4.1.
SAMPLE PRESSURE SENSOR
An absolute pressure transducer plumbed to the input of the analyzer’s sample chamber
is used to measure the pressure of the sample gas before it enters the chamber. This
upstream used to validate the critical flow condition (2:1 pressure ratio) through the
instrument’s critical flow orifice (refer to Section 9.4.2). Also, if the
Temperature/Pressure Compensation (TPC) feature is turned on (refer to Section 9.7.3),
the output of this sensor is also used to supply pressure data for that calculation.
The actual pressure measurement is viewable through the analyzer’s front panel display
as the test function
PRESS
.
9.4.4.2.
SAMPLE FLOW SENSOR
A thermal-mass flow sensor is used to measure the sample flow through the analyzer.
This sensor is also mounted on the pneumatic pressure/flow sensor board upstream of
the sample chamber. The flow rate is monitored by the CRT which issues a warning
message (
SAMP FLOW WARN
) if the flow rate is too high or too low.
The flow rate of the sample gas is viewable via the front panel as the
SAMP FL
test
function.
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...