Calibration Procedures
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
188
With dynamic calibration turned on, the state of the internal setup variables
DYN_SPAN
and
DYN_ZERO
is set to
ON
and the instrument will reset the slope and
offset values for the SO
2
response each time the AutoCal program runs. This continuous
re-adjustment of calibration parameters can often mask subtle fault conditions in the
analyzer. It is recommended that, if dynamic calibration is enabled, the analyzer’s test
functions, slope and offset values be checked frequently to assure high quality and
accurate data from the instrument.
6.9. CALIBRATION QUALITY
After completing one of the calibration procedures described above, it is important to
evaluate the analyzer’s calibration
SLOPE
and
OFFSET
parameters. These values
describe the linear response curve of the analyzer. The values for these terms, both
individually and relative to each other, indicate the quality of the calibration. To perform
this quality evaluation, you will need to record the values of both test functions (refer to
Section 4.2.1 or Appendix A-3), all of which are automatically stored in the DAS
channel
CALDAT
for data analysis, documentation and archival.
Ensure that these parameters are within the limits listed in the following Table.
Table 6-5: Calibration Data Quality Evaluation
FUNCTION
MINIMUM VALUE
OPTIMUM VALUE
MAXIMUM VALUE
SLOPE
-0.700
1.000
1.300
OFFS
50.0 mV
n/a
250.0 mV
These values should not be significantly different from the values recorded on the
Teledyne ML
Final Test and Validation Data
sheet that was shipped with your
instrument. If they are, refer to troubleshooting in Section 10.
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...