Teledyne ML - T100 UV Fluorescence SO2 Analyzer
Calibration Procedures
187
To program the sample sequence shown Table 6-4:
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL CALZ CZLS
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG
ACAL
DAS RNGE PASS CLK MORE EXIT
SETUP X.X
SEQ 1) DISABLED
NEXT
MODE
EXIT
SETUP X.X
SEQ 2) DISABLED
PREV NEXT
MODE
EXIT
SETUP X.X
MODE: DISABLED
NEXT
ENTR EXIT
SETUP X.X
MODE: ZERO
PREV
NEXT
ENTR EXIT
SETUP X.X
MODE: ZERO
–SPAN
PREV
NEXT
ENTR
EXIT
SETUP X.X
SEQ 2) ZERO
–SPAN, 1:00:00
PREV NEXT
MODE
SET
EXIT
SETUP X.X
TIMER ENABLE: ON
SET>
EDIT
EXIT
SETUP X.X
STARTING DATE: 01
–JAN–02
<SET SET>
EDIT
EXIT
SETUP X.X
STARTING DATE: 01
–JAN–02
0
4 SEP
0
3
ENTR
EXIT
Toggle keys
to set
day, month &
year:
Format :
DD-MON-Y
Y
SETUP X.X
STARTING DATE: 04
–SEP–03
<SET
SET>
EDIT
EXIT
Default
value is
ON
SETUP C.4
STARTING DATE: 04
–SEP–03
<SET
SET>
EDIT
EXIT
SETUP C.4
STARTING TIME:00:00
<SET SET>
EDIT
EXIT
SETUP C.4
STARTING TIME:00:00
1 4 : 1 5
ENTR
EXIT
Toggle keys to
set time:
Format : HH:MM
This is a 24 hr
clock .
PM hours are
13
– 24.
Example
2:15 PM = 14:15
Toggle keys
to set
number of
days
between
procedures
(1-367)
Toggle keys
to set
delay time for
each iteration
of the
sequence:
HH:MM
(0
– 24:00)
SETUP C.4
DELTA DAYS: 1
<SET SET>
EDIT
EXIT
SETUP C.4
DELTA DAYS: 1
0 0 2
ENTR
EXIT
SETUP C.4
DELTA DAYS:2
<SET
SET>
EDIT
EXIT
SETUP C.4
DELTA TIME00:00
<SET SET>
EDIT
EXIT
SETUP C.4
DELTA TIME: 00:00
0 0 :3 0
ENTR
EXIT
SETUP C.4
DELTA TIEM:00:30
<SET
SET>
EDIT
EXIT
SETUP C.4
STARTING TIME:14:15
<SET
SET>
EDIT
EXIT
EXIT
returns
to the
SETUP
Menu
SETUP C.4
DURATION:15.0 MINUTES
<SET SET>
EDIT
EXIT
SETUP C.4
DURATION 15.0MINUTES
3 0 .0
ENTR
EXIT
SETUP C.4
DURATION:30.0 MINUTES
<SET
SET>
EDIT
EXIT
Toggle keys
to set
duration for
each
iteration of
the
sequence:
Set in
Decimal
minutes
from
0.1
– 60.0
SETUP C.4
CALIBRATE: OFF
<SET SET>
EDIT
EXIT
SETUP C.4 CALIBRATE: OFF
ON
ENTR
EXIT
SETUP C.4
CALIBRATE: ON
<SET SET> EDIT
EXIT
Toggle key
between
Off and
ON
SETUP C.4
SEQ 2) ZERO
–SPAN, 2:00:30
PREV NEXT
MODE
SET
EXIT
Sequence
Delta Time
MODE Delta Days
Figure 6-12:
AUTO CAL
– User Defined Sequence
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...