Calibration Procedures
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
184
6.7. MANUAL CALIBRATION IN DUAL OR AUTO REPORTING
RANGE MODES
When the analyzer is in either Dual or Auto Range modes the user must run a separate
calibration procedure for each range. After pressing the
CAL
,
CALZ
or
CALS
buttons
the user is prompted for the range that is to be calibrated as seen in the
CALZ
example
below:
Continue Calibration as per
Standard Procedure
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL
CALZ
CALS
SETUP
ZERO CAL M
RANGE = 500.000 PPB SO2 =XXX.X
< TST TST >
ZERO
CONC
EXIT
WAIT 10
MINUTES
Or until the
reading
stabilizes and
the
ZERO
button
is displayed
SAMPLE
RANGE TO CAL: LOW
LOW
HIGH
ENTR
SETUP
SAMPLE
RANGE TO CAL: HIGH
LOW HIGH
ENTR
SETUP
Analyzer enters
ZERO CAL
Mode
Refer to Table 6-1 for
Z/S Valve States
during this operating
mode.
Figure 6-11:
Manual Calibration in Dual/Auto Reporting Range Modes
Once this selection is made, the calibration procedure continues as previously described
in Sections 7.2 through 7.6. The other range may be calibrated by starting over from the
main
SAMPLE
display.
6.7.1. CALIBRATION WITH REMOTE CONTACT CLOSURES
Contact closures for controlling calibration and calibration checks are located on the rear
panel
CONTROL IN
connector. Instructions for setup and use of these contacts can be
found in Section 5.3.1.2.
When the appropriate contacts are closed for at least 5 seconds, the instrument switches
into zero, low span or high span mode and the internal zero/span valves will be
automatically switched to the appropriate configuration. The remote calibration contact
closures may be activated in any order. It is recommended that contact closures remain
closed for at least 10 minutes to establish a reliable reading; the instrument will stay in
the selected mode for as long as the contacts remain closed.
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...