Calibration Procedures
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
176
M-P CAL
SO2 STB=X.XXX PPB
SO2
=XXX.X
< TST TST >
ENTR
CONC
EXIT
M-P CAL
SO2 STB=X.XXX PPB
SO2
=XXX.X
< TST TST >
CAL
SETUP
M-P CAL
SO2 STB=X.XXX PPB
SO2
=XXX.X
< TST TST >
ZERO
CONC
EXIT
EXIT
returns to the main
SAMPLE
display
ACTION:
Allow zero gas to enter the sample port at the
rear of the instrument.
M-P CAL
SO2 STB=X.XXX PPB
SO2
=XXX.X
< TST TST >
SPAN
CONC
EXIT
M-P CAL RANGE = 500.0 PPB
SO2
=XXX.X
< TST TST >
ENTR
SPAN CONC
EXIT
Wait until
SO2 STB
falls below
0.5 ppb.
This may take several
minutes.
The
SPAN
key now appears
during the transition from
zero to span.
You may see both keys.
SAMPLE RANGE = 500.0 PPB SO2 =XXX.X
< TST TST >
CAL
SETUP
Set the Display to show the
SO2 STB
test function.
This function calculates the
stability of the SO
2
measurement
Press
ENTR
to changes the
OFFSET
&
SLOPE
values for the
SO
2
measurements.
Press
EXIT
to leave the calibration
unchanged and return to the
previous menu.
ACTION:
Allow span gas to enter the sample port at the
rear of the instrument.
The value of
SO2 STB
may jump
significantly.
Wait until it falls back
below
0.5 ppb.
This may take several
minutes.
Press
ENTR to
change the
OFFSET
&
SLOPE
values for the
SO
2
measurements.
Press
EXIT
to leave the calibration
unchanged and return to the
previous menu.
SAMPLE RANGE = 500.0 PPB SO2 =XXX.X
< TST
TST
>
CAL
SETUP
M-P CAL
RANGE = 500.0 PPB
SO2
=XXX.X
< TST TST > ENTR CONC
EXIT
Figure 6-3: Setup for Manual Calibration without Z/S valve or IZS Option (Step 3)
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...