Remote Operation & Advanced Features
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
140
The Channel Properties, Triggering Events and Data Parameters/Functions for these
default channels are:
NAME:
PNUMTC
EVENT:
ATIMER
PARAMETERS:
2
STARTING DATE:
01-JAN-07
SAMPLE PERIOD:
000:00:05
REPORT PERIOD:
001:00:00
NO. OF RECORDS:
360
RS-232 REPORT:
OFF
COMPACT REPORT:
OFF
CHANNEL ENABLED:
ON
CAL HOLD OFF:
OFF
NAME:
CALDAT
EVENT:
SLPCHG
PARAMETERS:
3
NO. OF RECORDS:
200
RS-232 REPORT:
OFF
COMPACT REPORT:
OFF
CHANNEL ENABLED:
ON
CAL HOLD OFF:
OFF
PARAMETER:
SMPFLW
MODE:
AVG
PRECISION:
1
STORE NUM SAMPLES
OFF
PARAMETER:
SMPPRS
MODE:
AVG
PRECISION:
1
STORE NUM SAMPLES
OFF
PARAMETER:
OFSET1
MODE:
INST
PRECISION:
1
STORE NUM SAMPLES
OFF
PARAMETER:
SLOPE1
MODE:
INST
PRECISION:
3
STORE NUM SAMPLES
OFF
PARAMETER:
ZSCNC1
MODE:
INST
PRECISION:
1
STORE NUM SAMPLES
OFF
PARAMETER:
CONC1
MODE:
AVG
PRECISION:
1
STORE NUM SAMPLES
OFF
PARAMETER:
STABIL
MODE:
AVG
PRECISION:
2
STORE NUM SAMPLES
OFF
NAME:
CONC
EVENT:
ATIMER
PARAMETERS:
2
STARTING DATE:
01-JAN-07
SAMPLE PERIOD:
000:00:01
REPORT PERIOD:
000:00:05
NO. OF RECORDS:
4032
RS-232 REPORT:
ON
COMPACT REPORT:
OFF
CHANNEL ENABLED:
ON
CAL HOLD OFF:
ON
LIST OF CHANNELS
NAME:
DETAIL
EVENT:
ATIMER
PARAMETERS:
14
STARTING DATE:
01-JAN-07
SAMPLE PERIOD:
000:00:01
REPORT PERIOD:
000:01:00
NO. OF RECORDS:
480
RS-232 REPORT:
OFF
COMPACT REPORT:
OFF
CHANNEL ENABLED:
ON
CAL HOLD OFF:
OFF
NAME:
FAST
EVENT:
ATIMER
PARAMETERS:
14
STARTING DATE:
01-JAN-07
SAMPLE PERIOD:
000:00:01
REPORT PERIOD:
000:00:01
NO. OF RECORDS:
360
RS-232 REPORT:
OFF
COMPACT REPORT:
OFF
CHANNEL ENABLED:
ON
CAL HOLD OFF:
OFF
PARAMETER:
UVDET
MODE:
AVG
PRECISION:
4
STORE NUM SAMPLES
OFF
PARAMETER:
LAMPR
MODE:
AVG
PRECISION:
4
STORE NUM SAMPLES
OFF
PARAMETER:
DRKPMT
MODE:
AVG
PRECISION:
4
STORE NUM SAMPLES
OFF
PARAMETER:
DARKUV
MODE:
AVG
PRECISION:
4
STORE NUM SAMPLES
OFF
PARAMETER:
CONC1
MODE:
AVG
PRECISION:
3
STORE NUM SAMPLES
OFF
PARAMETER:
STABIL
MODE:
AVG
PRECISION:
4
STORE NUM SAMPLES
OFF
PARAMETER:
STRLGT
MODE:
AVG
PRECISION:
4
STORE NUM SAMPLES
OFF
PARAMETER:
RCTEMP
MODE:
AVG
PRECISION:
2
STORE NUM SAMPLES
OFF
PARAMETER:
SMPPRS
MODE:
AVG
PRECISION:
4
STORE NUM SAMPLES
OFF
PARAMETER:
BOXTEMP
MODE:
AVG
PRECISION:
4
STORE NUM SAMPLES
OFF
PARAMETER:
HVPS
MODE:
AVG
PRECISION:
4
STORE NUM SAMPLES
OFF
PARAMETER:
REFGND
MODE:
AVG
PRECISION:
4
STORE NUM SAMPLES
OFF
PARAMETER:
REF4096
MODE:
AVG
PRECISION:
4
STORE NUM SAMPLES
OFF
PARAMETER:
PMTDET
MODE:
AVG
PRECISION:
4
STORE NUM SAMPLES
OFF
Same parameters &
settings as
DETAIL
Figure 5-9: Default DAS Channels Setup
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...