Teledyne ML - T100 UV Fluorescence SO2 Analyzer
Operating Instructions
123
4.8.10. TEST CHANNEL OUTPUT
When activated, output channel
A3
can be used in the standard configuration to report
one of the test functions viewable from the SAMPLE mode display.
To activate the
A3
channel and select a test function, access the Signal I/O from the
DIAG Menu (refer to Figure 4-17), then press:
DIAG TCHN
TEST CHANNEL: NONE
NEXT
ENTR
EXIT
DIAG
TEST CHAN OUTPUT
PREV
NEXT
ENTR
EXIT
Continue to press
NEXT
until …
DIAG TCHN
TEST CHANNEL: PMT READING
PREV
NEXT
ENTR
EXIT
Press
PREV
or
NEXT
to move through the
list of available
parameters
(Table 6-9)
Press
EXIT
to
return to the
DIAG
menu
Press
ENTR
to
select the displayed
parameter and to
activate the test
channel.
DIAG
SIGNAL I / O
PREV
NEXT
ENTR
EXIT
Figure 4-36:
DIAG
– Test Channel Output
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...