SyriXus Series Pumps Installation and Operation Guide
Section 9 Pump Maintenance, Troubleshooting, and Servicing
9-4
EAR99 Technology Subject to Restrictions Contained on the Cover Page
9.5 Resetting the
System
This action
completely deletes
user programmed settings, erasing
all programs and returning the controller to factory default settings.
Record your program settings and parameters before performing a
reset operation. If
ZERO
PRESS
has been employed, all corrected offsets
will be lost.
9.5.1 Basic Reset
To perform a
basic reset
, press
MENU
>
SYSTEM
RESET
(5) >
CONTINUE
(A).
9.5.2 Hard Reset
To perform a
hard reset
:
1. Turn the controller switch to
STANDBY
.
2. Press and hold
CLEAR
ENTRY
.
3. While still holding
CLEAR
ENTRY
, turn the controller switch to
ON
.
Continue holding down
CLEAR
ENTRY
for one second.
4. Release the
CLEAR
ENTRY
key and turn the controller switch back
to
STANDBY
.
5. Turn the controller switch back to ON, and perform the
Basic
Reset
, as described above.
9.6 Diagnostic Menu
The syringe pump controller can run a number of system tests that
are initialized using the keypad.
Figure 9-1
Accessing the diagnostic menu
9.6.1 LCD TEST
Cycles all segments of the display through all displayed characters. To
stop the test, press any key.
1. LCD TEST
2. ANALOG INPUTS
3. A/D CAL
4. KEYPAD TEST
5. MOTOR CONTROL/LIMITS
6. SERIAL TEST
PREVIOUS
MORE
(
MORE
)
(
DIAGNOSTIC
MENU
)
(
MORE
)