Running Verification Script Engine (VSE)
278
Sierra M6
‐
2 SAS/SATA Protocol Analyzer User Manual
3. Select a Compliance Type: Device or Host
4. Select a Device Type: ATA or ATAPI
5. Enter a value for the
Number
of
Runs
to run a particular test more than once.
6. Enter the
Capture
Timeout
value in seconds
.
7. To view failed test traces in the sample viewer, check the
Automatically load failed test traces
into
Sierra
trace
viewer
check box.
8. To save the failed test traces only, check the
Save
only
failed
test
traces
check box.
9. Enter a path and file name for
Report
file
, or use the default file name and path:
For Windows XP:
C:\Program Files\LeCroy\SAS/SATA Protocol
Suite\User\Compliance_Device1.rtf
For Windows 7:
C:\Users\Public\Documents\LeCroy\SAS SATA Protocol
Suite\User\Compliance_Device1.rtf
10. Enter a path and file name for
Saved
Traces
, or use the default folder:
For Windows XP:
C:\Program Files\LeCroy\SAS SATA Protocol Suite\User\
For Windows 7:
C:\Users\Public\Documents\LeCroy\SAS SATA Protocol Suite\User\
11. After you select tests, click
Start
.
12. The test runs and, after a brief period, displays the result.
13. To save the current compliance setup for later use, click the
Save
button to open the
Save As dialog. Assign a meaningful name to the setup and save it as a
*.cst
compliance file.
14. To run a previously defined setup, click the
Load
button and choose a previously
defined setup to run.
Note:
When you are running the Compliance Test, you cannot work with Trainer and Emulator while
the device is connected to a D connector and HBA is connected to an H connector.
3.7
Running Verification Script Engine (VSE)
You can perform custom post
‐
process analysis of the open trace by running a verification
script over the trace. A verification script instructs the application to send trace and
analysis information to the script. A verification script also contains script code, written
using CATC Script Language (CSL) (see the
CSL_RefManual.pdf
document in the
docs
directory of the installation), used to process trace data and output that data in different
formats.
Summary of Contents for Sierra M6-2
Page 37: ...Sierra M6 2 SAS SATA Protocol Analyzer User Manual 35 Expandability Teledyne LeCroy ...
Page 375: ...Sierra M6 2 SAS SATA Protocol Analyzer User Manual 373 Interface Teledyne LeCroy ...
Page 436: ...Teledyne LeCroy Scenario Batch Files 434 Sierra M6 2 SAS SATA Protocol Analyzer User Manual ...
Page 526: ...Teledyne LeCroy 524 Sierra M6 2 SAS SATA Protocol Analyzer User Manual ...