Message
Description
Action
Failed to set up DMA Card 0,
ERROR =
Failed to allocate memory for
ADMXRC transfer.
Failed to set ADMXRC clock,
Card_1, ERROR =
Failed to get ADMXRC space info,
Card_1, ERROR =
Failed to configure 1st FPGA with
<x>, ERROR =
Failed to find ADMXRC info,
card_0, ERROR =
Failed to set ADMXRC clock,
Card_0, ERROR =
Failed to get ADMXRC space info,
Card_0, ERROR =
Failed to find ADMXRC info,
card_1, ERROR =
A.2.3 Data Reading Errors
Message
Description
Action
DMA Transfer Timeout. <x> of <x>
bytes received
This error generally indicates
that the uplink from the
receiver is interrupted at
some point.
Check the BITE tab for
receiver BITE values to
try and localize the
failure to a component.
Check uplink cable
connections. Check
temperature of the SSP
via BITE. Restore
factory defaults. Reboot
the PC.
FATAL: DMA Read size limitation
reached.
Lost sync with DMA read
A general error occurred
while transferring data to the
software.
Reboot the system.
A.2.4 System Resource Errors
Message
Description
Action
Error starting <name> thread.
A module has failed to start
correctly.
Reboot the system.