PI-MTE System Manual
Issue 6
7.7.1
Change the DMA Buffer Setting
Perform the following procedure to change the DMA Buffer size configuration:
1. Click OK to dismiss the error dialog.
2. Reboot WinView.
3. Note the array size displayed in the Full Chip dimension field on the
Setup
►
Hardware
►
Controller/CCD tab or the Acquisition
►
Experiment
Setup
►
Main tab.
4. Open Setup
►
Environment
►
Environment dialog box.
5. Increase the DMA buffer size to a minimum of:
•
32 MB;
•
64 MB if it is currently 32 MB;
•
128 MB if it is currently 64 Mb.
6. Click OK and then close WinView.
7. Reboot the host computer.
8. Launch WinView and begin acquiring data or focusing.
NOTE:
If this message is displayed again, increase the DMA buffer
size again.
7.8
Data Overrun Occurred
Because of memory constraints and the way that USB transfers data, a Data overrun has
occurred message may be displayed during data acquisition. If this message is
displayed, perform one or more of the following actions:
•
Minimize the number of programs running in the background while you are
acquiring data with WinView/32.
•
Run data acquisition in Safe Mode.
•
Add memory.
•
Use binning.
•
Increase the exposure time.
•
Defragment the hard disk.
•
Update the Orange Micro USB2 driver.
Refer to
Section 4.6.2, Update the OrangeUSB USB 2.0 Driver
complete information.
If the problem persists, your application may be USB 2.0 bus limited. Since the host
computer controls the USB 2.0 bus, there may be situations where the host computer
interrupts the USB 2.0 port. In most cases, you will not notice the interrupt. However,
there are some instances when the data overrun cannot be overcome because USB 2.0
bus limitations combined with long data acquisition times and/or large data sets
increase the possibility of an interrupt while data is being acquired. If your experiment
requirements include long data acquisition times and/or large data sets, your
application may not be suitable for the USB 2.0 interface. If this is not the case and data
overruns continue to occur, contact Customer Support. Refer to
on page 116 for complete information.
Summary of Contents for Princeton Instruments PI-MTE
Page 1: ...www princetoninstruments com PI MTE System Manual 4411 0097 Issue 6 October 1 2019...
Page 22: ...22 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 25: ...Chapter 3 Installation 25 PI MTE System Diagram 4411 0097_0014...
Page 26: ...26 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 58: ...58 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 90: ...90 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 94: ...94 PI MTE System Manual Issue 6 Figure B 2 PI MTE Camera 8 01 05 and later 4411 0097_0054...
Page 95: ...Appendix B Outline Drawings 95 Figure B 3 PI MTE Camera 8 01 05 and earlier 4411 0097_0055...
Page 102: ...102 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 112: ...112 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 117: ...This page is intentionally blank...