Chapter 4
System Setup
29
•
Inlet/Outlet Port Locations
VCR male fittings (or Swagelok fittings for earlier systems) and flexible stainless
steel hoses are provided to make the hose connections between the camera
and an intermediate vacuum flange. The ports are not interchangeable.
•
Coolant Temperature
+10°C to +15°C, typical.
WARNING!
If a lower temperature is set, there can be condensation on
the external tubing or if the camera is being used outside of
a vacuum chamber, condensation may occur on or inside
the camera. Damage due to condensation may not be
covered by the warranty.
4.3.4
Power
•
Camera
The PI-MTE camera receives its power from the controller, which in turn plugs
into a source of AC power.
•
ST-133:
The ST-133 Controller can operate from any one of four different nominal line
voltages: 100, 120, 220, or 240 V AC. Refer to the Fuse/Voltage label on the
back of the ST-133 for fuse, voltage, and power consumption information.
The plug on the power cord supplied with the system should be compatible
with the line-voltage outlets in common use in the region to which the system
is shipped. If the power cord plug is incompatible, a compatible plug should be
installed, taking care to maintain the proper polarity to protect the equipment
and assure user safety.
4.3.5
Host Computer
NOTE:
Computers and operating systems all undergo frequent
revision. The following information is only intended to give
an approximate indication of the computer requirements.
Please contact the factory to determine your specific needs.
The host computer must meet or exceed the following requirements:
•
PC with Pentium 3 or better processor and runs at 1 GHz or better.
•
Windows
®
XP (with Service Pack 1), Windows Vista
®
(32-bit), or Windows 7
(32-bit) operating system.
•
Native USB 2.0 support on the motherboard or USB Interface Card (Orange
Micro 70USB90011 USB2.0 PCI is recommended for desktop computers and the
SIIG, Inc. USB 2.0 PC Card, Model US2246 is recommended for laptop
computers).
•
Minimum of 256 MB of RAM.
•
CD-ROM drive.
•
Hard disk with a minimum of 80 MB available. A complete installation of the
program files takes about 17 MB and the remainder is required for data storage,
depending on the number and size of images or spectra collected. Disk level
compression programs are not recommended.
Summary of Contents for Princeton Instruments PI-MTE
Page 1: ...www princetoninstruments com PI MTE System Manual 4411 0097 Issue 6 October 1 2019...
Page 22: ...22 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 25: ...Chapter 3 Installation 25 PI MTE System Diagram 4411 0097_0014...
Page 26: ...26 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 58: ...58 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 90: ...90 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 94: ...94 PI MTE System Manual Issue 6 Figure B 2 PI MTE Camera 8 01 05 and later 4411 0097_0054...
Page 95: ...Appendix B Outline Drawings 95 Figure B 3 PI MTE Camera 8 01 05 and earlier 4411 0097_0055...
Page 102: ...102 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 112: ...112 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 117: ...This page is intentionally blank...