39
Chapter 5: System Setup
This chapter provides information and procedures necessary to setup a PI-MAX3
system for both imaging and spectroscopic applications.
WARNING!
Voltages inside the PI-MAX3 may exceed 6,000 volts. To
avoid possible hazard to personnel, use the instrument only
according to the directions in this manual and only for the
purposes for which it is designed. Never attempt to operate
the PI-MAX3 with its covers removed.
CAUTION!
!
Image intensified detectors can be destroyed if exposed to
excessive light levels. Teledyne Princeton Instruments
cannot take responsibility for PI-MAX3 detector damage
due to misuse.
Intensified cameras are particularly susceptible to overload
damage when operated in conjunction with high-intensity
light sources such as lasers. Spot damage can occur without
the overload condition being detected.
NOTE:
Appendix A, Technical Specifications
, on page 197
for information about system requirements, including
environmental, storage, power, and minimum host
computer requirements.
The PI-MAX3 has an audible alarm and shutdown circuit to bias the photocathode off if
excessive photocathode current is detected. The circuit automatically resets and biases
the photocathode back on after approximately 0.5 seconds. The short-term protection
provided will not prevent intensifier damage if excessive light is allowed to continuously
fall on the intensifier.
It is also possible for excessively bright spots to damage the intensifier tube without
triggering the alarm.
5.1
Unpack the System
During unpacking, check the system components for possible signs of shipping
damage. If there are any, notify Teledyne Princeton Instruments and file a claim with the
carrier. Be sure to save the shipping carton for inspection by the carrier. If damage is not
apparent but system specifications cannot be achieved, internal damage may have
occurred in shipment.
Save the original packing materials so you can safely ship the camera system to another
location or return it to Teledyne Princeton Instruments for repairs if necessary.
Summary of Contents for Princeton Instruments PI-MAX 3
Page 1: ...www princetoninstruments com PI MAX 3 System Manual 4411 0129 Issue 2 August 20 2019 ...
Page 32: ...32 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 38: ...38 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 124: ...124 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 150: ...150 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 174: ...174 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 178: ...178 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 190: ...190 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 214: ...214 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 236: ...236 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 244: ...244 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 249: ...This page is intentionally blank ...