![Teledyne Princeton Instruments PI-MAX 3 System Manual Download Page 241](http://html.mh-extra.com/html/teledyne/princeton-instruments-pi-max-3/princeton-instruments-pi-max-3_system-manual_1080556241.webp)
Appendix H
Glossary
241
Intensifier Gating
Speed:
Temporal resolution in a PI-MAX3 is made possible by switching
the intensifier on and off (gating) very rapidly. Typical fast-gate
intensifiers have minimum gate widths (FWHM) of approximately
2 ns. For slow-gated devices the FWHM is about 50 ns. Refer to
Fast-gating is achieved by adding a nickel (Ni) underlayer to
photocathode. However, this layer may produce an effective QE
reduction of as much as 40%. Slow-gate intensifiers have neither
the Ni layer nor its effects on QE.
Intensifier
On/Off Ratio:
The ratio of light output when the intensifier is gated on and off:
The higher the ratio, the better the gating. A high on/off ratio is
necessary to eliminate the background and to faithfully
reproduce transient events. In the visible region on/off ratios
exceeding 10
6
:1 is typically achieved. In the UV region, the on/off
ratio is typically much poorer (10
4
:1) though with MCP Bracket
Pulsing ratios in the UV region can be improved dramatically
(10
7
Section 8.3, MCP Bracket Pulsing
Intensifier Size:
18 mm diameter and 25 mm diameter. Generally speaking, the
larger diameter gives a larger field of view at the surface of the
CCD array. The coupling of the intensifier to the CCD array is also
a factor in determining the field view. A fiberoptic reducing taper
of 1.27:1 will increase the field of view, while a taper of 1:1 will
have no effect.
Intensifier Types:
Refer to:
•
•
;
•
;
•
•
;
•
.
lp/mm:
line pairs per millimeter. A measure of resolution based on the
ability of the imaging system to differentiate between two parallel
lines. The higher the value, the finer the resolution.
lux:
The SI unit of illuminance, equal to one lumen per square meter.
MCP:
MicroChannel Plate.
Composed of cylindrical channels through which electrons from
the photocathode travel and generate additional electrons,
resulting in electron gain. At the output of the MCP is a
phosphor-coated fluorescent screen that converts the electrons
to photons that subsequently strike the CCD array and generate
charge in the array pixels.
Summary of Contents for Princeton Instruments PI-MAX 3
Page 1: ...www princetoninstruments com PI MAX 3 System Manual 4411 0129 Issue 2 August 20 2019 ...
Page 32: ...32 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 38: ...38 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 124: ...124 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 150: ...150 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 174: ...174 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 178: ...178 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 190: ...190 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 214: ...214 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 236: ...236 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 244: ...244 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Page 249: ...This page is intentionally blank ...