8. Verify there are no leaks or air bubbles in the hoses.
NOTE:
Small air bubbles (approximately the size of bubbles in soda)
are common, particularly immediately following start up.
These bubbles do not prevent proper operation.
•
If no problems are observed, proceed to step 9.
•
If there are leaks or air bubbles, turn the circulator off and correct the
problem(s) by securing the hoses or adding more coolant to the reservoir.
Turn the circulator back on. Recheck and if there are no problems, proceed to
step 9.
9. Turn on the camera.
10. Launch LightField.
5.3.1
Experiment Shutdown
Following the completion of a liquid-cooled experiment, perform the following
procedure:
1. Turn off the Lansis camera’s power supply.
2. Turn off the chiller according to all manufacturer-supplied documentation and
procedures.
3. If desired, carefully disconnect the coolant hoses from the Lansis camera.
WARNING!
If the Lansis is to be shipped, to avoid potential catastrophic
damage to the camera, all coolant must be drained from it.
Refer to