HDO9000 High Definition Oscilloscopes Operator's Manual
Maintenance
Topics in this section describe procedures for keeping the instrument in optimal working condition.
Touch Screen Calibration
Periodically calibrate the touch screen to maintain its accuracy and responsiveness. We recommend
that you use a stylus rather than your finger for this procedure.
1. From the menu bar, choose
Utilities > Utilities Setup
.
2. On the Utilities main dialog, touch
Touch-Screen Calibration
.
3. Following the prompts, touch as close as possible to the center of each cross that appears on
the screen until the calibration sequence is complete.
Restart/Reboot Instrument
To restart the oscilloscope application, choose
File > Exit
from the menu bar, then touch the
Start
DSO
desktop shortcut.
Note:
You will generally need to exit and restart the application after adding new options keys.
To reboot the instrument, which includes restarting the OS:
1. From within the oscilloscope application, choose
File > Shutdown
from the menu bar. From the
Windows desktop, use the Windows Start Menu Shutdown command.
2. Wait 10 seconds, then press the
Power button
on the front of the instrument.
164