5-8
OPERATOR MAINTENANCE
FIRST RESPONDER CG - USER HANDBOOK
Water Filter (Accessory)
To replace the Hydrophobic Filter in the Water Filter
assembly, proceed as follows:
1.
Unscrew the filter assembly (Fig. 5-8).
Water Filter Housing
Assembly
Hydrophobic Filter
Assembly
Probe Washer
Fig. 5-8 Water Filter Assembly
2.
Remove Hydrophobic Filter, then discard.
3. Check Filter assembly Probe Washer for signs of
damage or wear and replace if necessary.
4. Fit new Hydrophobic Filter.
Note: The filter should be installed with the yellow label
on the filter, facing the yellow label on the housing
flange.
5. Reassemble the Filter assembly.