Falcon GC
Operation
Teledyne Analytical Instruments
17
Operation
4.1 Starting the GC
4.1.1 Turning on the GC Modules
Now that the method file has been downloaded to the GC you are
ready to activate the Falcon modules and heat the instrument.
From the home screen on the Falcon touch screen, simply touch each
module icon individually and turn on their temperature and pressure
controllers. Press “Back” to return to the home screen from each
module. The order is not important as long as you turn on the injector
oven module and carrier gas first. Conversely, you can also access each
module by pressing Menu→Modules.
From the home screen you should be able to monitor the current
temperature or pressure of each module. You will know the setpoint has
been reached when the indicator next to the specific temperature or
pressure turns green. Yellow is below the setpoint and red is over
setpoint.
Summary of Contents for FALCON GC
Page 10: ...Falcon GC x Teledyne Analytical Instruments Blank Page ...
Page 15: ...Falcon GC Preparatio n Teledyne Analytical Instruments 5 3 Click on Report ...
Page 16: ...Preparation Falcon GC 6 Teledyne Analytical Instruments Blank Page ...
Page 42: ...Maintenance Troubleshooting Falcon GC 32 Teledyne Analytical Instruments ...
Page 62: ...Appendix Falcon GC 52 Teledyne Analytical Instruments A 3 Drawings ...
Page 63: ...Falcon GC Appendix Teledyne Analytical Instruments 53 ...
Page 64: ...Appendix Falcon GC 54 Teledyne Analytical Instruments ...
Page 65: ...Falcon GC Appendix Teledyne Analytical Instruments 55 ...
Page 66: ...Appendix Falcon GC 56 Teledyne Analytical Instruments ...
Page 67: ...Falcon GC Appendix Teledyne Analytical Instruments 57 ...