Programming Manual
–
T3AWG3K-C Series Arbitrary Waveform Generator
True-Arb Operating Mode
64
the sampling frequency), synchronous with the analog outputs, for each se-
quencer entry and for each repetition.
MARKer:MODE1?
Might return FIXEDHigh
Table 69: MARKer:MODE[m]
Command
MARKer:SKEW[m]
Description
This command and query sets or returns the marker delay.
It defines the skew between the marker and the analog channels.
The edited value is automatically rounded to the closest value that
the hardware can implement.
Group
Marker
Syntax
MARKer:SKEW[m] {MINimum|MAXimum|DEFault|<Seconds>}
MARKer:SKEW[m]? [{MINimum|MAXimum}]
Important Note:
The maximum value you can set in the marker skew
depends on the Sampling rate.
Related Commands
None
Arguments
•
MINimum: sets the parameter to the minimum value
•
MAXimum: sets the parameter to the maximum value
•
DEFault: sets the parameter to the default value
•
< Volts > := <NRf> value
The value of m indicates the Marker Out number.
Returns
<NRf>
Example
MARKer:SKEW1 1E-9
Sets the marker delay to 1ns
MARKer:SKEW1?
Might returns “1E
-
3”
Table 70: MARKer:SKEW[m]