BLZZRD Portable Refrigerated Sampler
Section 7 SDI-12 Sondes
7-7
7.6 Sonde Setup - Storing
Parameter Data
Whether you are using an ISCO-ready sonde or other SDI-12
compatible sonde, you will need to tell the sampler controller if
you want to store all of the parameter data or data from selected
parameters.
Data is stored in memory partitions. A fixed amount of memory
is available to store data. If you are using some parameters to
simply trigger an event or enable condition, it is not necessary to
store the data.
If
all
of the parameters’ data will fit in memory, you will be asked
if you want to store all of the parameter data. Answering NO will
still allow you to selectively store parameter data; otherwise, all
data will be stored.
You will be asked to store parameter data during the setup
process. You can also manually change the setup using options in
the Hardware menu.
To access the setup option in the Hardware menu, select Other
Function from the main menu, then choose Hardware. Advance
through the screens until you see one that displays SDI-12
DATA. Select this screen to display the following:
Select SETUP KNOWN SONDES and then indicate the address
of the sonde you want to setup. A screen containing the sonde
model and number of data values will appear briefly. Then you
will be asked to verify, parameter by parameter, what data you
want to store. The screens will look similar to the one below:
Indicate if you want to store the data for that particular
parameter and then advance through the rest of the parameter
screens. Once you have defined your data storage options, select
DONE.
You will be prompted that you have entered a new hardware
setup and asked if you want to erase the stored data. Indicate
YES or NO. A partition memory status message will be displayed
and (if a parameter exists that can be calibrated) you will then
advance to the calibration screen (refer to section 7.7). If you do
not want to calibrate or validate, select DONE to return to the
Hardware Options screens.
NO SDI-12 SONDE
SETUP KNOWN SONDES
SCAN SDI-12 SONDES
SONDE 7, STORE
TEMPERATURE DATA?
YES NO
LABEL IS: TEMP7
Summary of Contents for Everywhereyoulook BLZZRD
Page 2: ......
Page 4: ......
Page 8: ...BLZZRD Portable Refrigerated Sampler Safety vi ...
Page 16: ...BLZZRD Portable Refrigerated Sampler Table of Contents xiv ...
Page 34: ...BLZZRD Portable Refrigerated Sampler Section 1 Introduction 1 18 ...
Page 58: ...BLZZRD Portable Refrigerated Sampler Section 2 Installation Preparation 2 24 ...
Page 72: ...BLZZRD Portable Refrigerated Sampler Section 3 Programming Introduction 3 14 ...
Page 88: ...BLZZRD Portable Refrigerated Sampler Section 4 Standard Programming 4 16 ...
Page 219: ...BLZZRD Portable Refrigerated Sampler Appendix B Material Safety Data Sheets B 8 ...
Page 233: ...BLZZRD Portable Refrigerated Sampler Appendix D Replacement Parts D 6 ...
Page 241: ...BLZZRD Portable Refrigerated Sampler Appendix F Battery Selection Guide F 4 ...