background image

BLZZRD Portable Refrigerated Sampler

Section 5  Extended Programming

5-17

Switch times occur regardless of the state of the sampler. Be

aware that the sampler may leave empty bottles if it becomes dis-

abled. Pause and resumes are an exception; the sampler will not

switch bottles during a pause if the current bottle is empty.
If the program settings instruct the sampler to take a sample at

the start time, the sampler also draws a sample at the switch

time. When this occurs, the pacing will also be reset at the switch

time.
Time switching is normally used with flow-paced programs that

do not take a sample at start. These programs will place a

sample in the current bottle if it is empty at the switch time, and

then move to the next bottle/set. Pacing intervals are reset at

every switch time.
When you use the time switching feature, the volume in each

bottle or bottle set may vary. Sample volumes that would exceed

the bottle capacity are not taken and a “PROBABLE OVERFILL”

message is logged.

5.9.7 Continuous Sampling

Sample programs can be run indefinitely by selecting “YES” at

the “RUN CONTINUOUSLY?” screen. Continuous sampling

resets the distribution when the distribution sequence is com-

plete. That is, when the last bottle of the set is reached, the next

bottle is the first bottle of the set. All pacing modes except

RANDOM are supported.
When running a program continuously, the bottles must be ser-

viced at regular intervals to prevent overfilling the bottles. The

sampler assumes that the next bottle/set is empty and ready to

receive samples. If you are using this feature with two-part pro-

gramming, keep in mind that the parts will most likely reset at

different intervals.

Note

Sampling and Combined Results reports are cleared when the
distribution sequence resets. If reports are needed, they must
be collected before sampler advances to the next bottle set.

Example: Time Switching Distribution

1. Enter  the  number  of  bottles  that  should  receive

samples at each sample event. 

2. Select TIME. 

3. Enter the time between bottle switches. 

4. Enter the first bottle switch time. 

 

2

 BOTTLES PER 

SAMPLE EVENT (1-max)

SWITCH BOTTLES ON:

NUMBER OF SAMPLES

TIME

SWITCH BOTTLES EVERY
__ HOURS, __ MINUTES

FIRST SWITCH TIME

AT: __:__

Summary of Contents for Everywhereyoulook BLZZRD

Page 1: ...973 063 Revision A March 2022 Copyright 2022 All rights reserved Teledyne ISCO Use and Disclosure of Data Information contained herein is classified as EAR99 under the U S Export Administration Regulations Export reexport or diversion contrary to U S law is prohibited ...

Page 2: ......

Page 3: ...n RMA specified Be sure to include a note describing the malfunction This will aid in the prompt repair and return of the equipment No item may be returned for service without a Return Material Authorization RMA number issued by Teledyne Teledyne ISCO welcomes suggestions that would improve the information presented in this manual or enhance the operation of the equipment itself Teledyne ISCO is c...

Page 4: ......

Page 5: ...3 Hz 2 Ampere power source using the attached AC line cord The internal AC power supply is protected by an internal 4A 250V input fuse Additionally the controller circuitry is protected by an internal 5 0 ampere PTC Positive Temperature Coeffi cient device WARNING Electrocution hazard Never manipulate electrical switches or power connections with wet hands or when your feet are in contact with wat...

Page 6: ...nstrument may be impaired this will increase your risk of injury AVERTISSEMENT Éviter les usages périlleux Si vous utilisez cet instrument d une manière autre que celles qui sont specifiées dans ce manuel la protection fournie de l instrument peut être affaiblie cela augmentera votre risque de blessure Additional safety information can be found in Appendix C Hazard Severity Levels This manual appl...

Page 7: ...isque de pincement Ces symboles vous avertit que les mains ou les doigts seront blessés sérieusement si vous les mettez entre les éléments en mouvement du mécanisme près de ces symboles Warnungen und Vorsichtshinweise Das Ausrufezeichen in Dreieck ist ein Warnzeichen das Sie darauf auf merksam macht daß wichtige Anleitungen zu diesem Handbuch gehören Der gepfeilte Blitz im Dreieck ist ein Warnzeic...

Page 8: ...BLZZRD Portable Refrigerated Sampler Safety vi ...

Page 9: ... Installing the 14 Bottle Kit 2 5 2 5 2 Installing the 4 Bottle Kit 2 8 2 5 3 Installing 2 5 Gallon Composite Bottle Kits 2 10 2 5 4 Installing the 5 Gallon Composite Bottle Kit 2 13 2 6 Installing a Power Source 2 14 2 6 1 Installing the Battery Box 2 14 2 6 2 AC Power Sources 2 15 2 6 3 12 VDC Power Sources 2 15 2 6 4 Battery Recommendations 2 15 2 7 Attaching the Suction Line 2 16 2 7 1 Attachi...

Page 10: ...nguage and Units of Length 4 1 4 3 Programming Examples 4 2 4 4 Pacing 4 5 4 4 1 Time Pacing 4 5 4 4 2 Flow Pacing 4 5 4 4 3 Trigger Pacing 4 6 4 5 Distribution 4 6 4 5 1 Sequential 4 8 4 5 2 Bottles Per Sample 4 8 4 5 3 Samples Per Bottle 4 8 4 5 4 Composite 4 8 4 5 5 Continuous Sampling 4 8 4 6 Start Times 4 8 4 6 1 How Do Start Times Work 4 9 4 6 2 Sampler Enable Disable 4 10 4 7 Programming fo...

Page 11: ...gramming Style 5 29 5 15 3 Measurement Interval 5 29 5 15 4 Dual Sampler Mode 5 29 5 15 5 Bottle Full Detection 5 30 5 15 6 Display Backlighting 5 30 5 15 7 Event Marks 5 30 5 15 8 Pre sample and Post sample Purge Counts 5 31 5 15 9 Serial Output 5 32 5 15 10 Interrogator Connector Power 5 34 5 16 Hardware Setup 5 35 5 16 1 SDI 12 Sonde Setup 5 35 5 16 2 Rain Gauge Setup 5 35 5 16 3 I O Pin Progra...

Page 12: ...7 7 7 7 Sonde Calibration and Validation 7 8 7 7 1 Calibration Screens 7 8 7 7 2 Validation Screen 7 9 7 8 Disconnecting Sondes 7 9 Section 8 Remote Operation 8 1 Computer Operation 8 1 8 1 1 Menu Control 8 2 8 1 2 External Program Control 8 4 8 1 3 Remote Control of Sampler Keypad 8 8 8 2 Telephone Commands 8 9 Section 9 Maintenance 9 1 Maintenance Checklist 9 1 9 2 Cleaning Guidelines 9 1 9 2 1 ...

Page 13: ...Determining the Power Consumption F 1 List of Figures 1 1 Software Update window 1 5 2 1 14 Bottle kit 2 6 2 2 Attaching the distributor arm and discharge tube 2 6 2 3 14 Bottle adapter plate installed 2 7 2 4 Carrier alignment 2 7 2 5 14 Bottle configuration completed 2 7 2 6 4 Bottle kit 2 8 2 7 4 Bottle adapter plate installed 2 9 2 8 4 Bottle configuration completed 2 9 2 9 Composite 2 5 gallo...

Page 14: ... 9 15 A 1 Menu Tree for Standard Programming A 1 A 2 Standard Programming Programming Screens A 2 A 3 Standard and Extended Programming Start Times A 3 A 4 Standard Programming Quick View Programming Screens A 4 A 5 Standard Programming Quick View Start Times A 5 A 6 View Report A 6 A 7 Maintenance Screens A 7 A 8 Manual Functions Screens A 8 A 9 Extended Programming Programming Screens A 9 A 10 E...

Page 15: ...3 5 Setting the Clock and Calendar 3 6 Entering a Site Description 3 8 Viewing Help Notes 3 9 Standard Program Sample every 15 Minutes One Sample per Bottle 4 2 Standard Program Flow Paced Sampling Two Bottles Per Sample 4 4 Start Times 4 9 Programming with 730 Bubbler Flow Module Installed 4 12 Programming with 750 Area Velocity Module Installed 4 14 Selecting a Stored Extended Program 5 3 Storm ...

Page 16: ...BLZZRD Portable Refrigerated Sampler Table of Contents xiv ...

Page 17: ...d samples dial out an alarm notification on its optional modem and then preserve the samples until they are retrieved For even greater versatility the sampler runs on either 12 VDC or AC line voltage The BLZZRD controller offers all of the control expansion and data management options available with Teledyne ISCO s 6700 Series Samplers including Plug in flow measurement modules eliminate the need ...

Page 18: ...ules The bay on the controller s side accepts any of Teledyne ISCO s 700 Series Modules The 700 Series includes 701 pH Parameter Module for monitoring pH and temperature 720 Submerged Probe Module for monitoring a flow stream s level and flow rate with a submerged probe 730 Bubbler Flow Module for monitoring a flow stream s level and flow rate with a bubbler system 750 Area Velocity Module for mon...

Page 19: ... Switched Data CSD communication is no longer available from Teledyne ISCO Samplers still using this modem must have software version 2 34 installed Refer to Section 1 4 5 for details 1 4 4 6712Ci Modem The 6712Ci is a CDMA cellular modem for serial communication via the internet This modem requires a customer supplied static IP address service contract with Verizon Refer to Section 1 4 5 for deta...

Page 20: ...l down to find the 6712 BLZZRD section 4 Select and download the appropriate language version you require 5 Rev D 6712 BLZZRD units require version 3 5 4 or above 6 Save the zip file to a directory of your choice and extract the contents Installing 2 50 or later Software CAUTION Updating the sampler s software will reset the program and memory partition settings to factory default settings and wil...

Page 21: ... information can be processed on a personal computer Readings and reports may be collected with a computer running Teledyne ISCO s Flowlink software Flowlink can download the information through a direct connection a modem connection when the sampler is equipped with the optional dial out modem or con nected to the optional digital cellular modem or from a 581 Rapid Transfer Device RTD The RTD is ...

Page 22: ...p Cover Protects pump and controller Controller Control panel sloped 15 degrees for easy reading Keys labeled with large vivid icons 80 character display 4 lines by 20 columns Display has selectable backlight Always on or always off Timed switching off when keypad is inactive for 60 seconds Memory for program and data storage Flash memory for easy software upgrades NEMA 4X and 6 IP67 ratings Sampl...

Page 23: ...low rate velocity pH temperature or the data collected by the 4 20 mA module Rain gauge and refrigerator temperature readings SDI 12 sonde readings Five Reports Available The Program Settings report listing current program settings The Sampling Results report listing the events occurring during the program The Combined Results report combining sampling events with readings from a rain gauge module...

Page 24: ...ite Samples deposited in a single large bottle Sequential Only one sample placed in each bottle Samples Per Bottle Multiple samples placed in each bottle Bottles Per Sample One sample deposited in multiple bottles Three Flexible Start Time Settings Start Immediately Starts the sampling program immediately Delayed Start Starts the sampling program after a user definable delay of 1 to 999 minutes Cl...

Page 25: ... the flow stream to the liquid detec tor Extended programming has two settings Auto Suction Head The head is automatically determined Fixed Suction Head A user definable measurement for the head Suction Line Rinses Program setting for the number of times 0 to 3 that the BLZZRD rinses the suc tion line before drawing a sample Sampling Retries Program setting for the number of times 0 to 3 that the ...

Page 26: ... pressure between 29 92 and 31 89 inHg 760 and 810 mmHg and a power source of a Teledyne ISCO High Capacity Power Pack at 12 5 volts DC no load output Controller Pump and Tubing Mechanical Specifications Weight of Controller Controller only 13 0 lbs 5 9 kg With pump tube 13 2 lbs 6 0 kg Controller Dimensions Length 10 3 in 26 0 cm Width 12 5 in 31 7 cm Height 10 in 25 4 cm Temperature Ranges Opera...

Page 27: ...minute 2 5 ft s 0 77 m s 23 feet 3 1 liters per minute 2 3 ft s 0 71 m s 25 feet 2 8 liters per minute 2 2 ft s 0 66 m s Maximum Suction Head 28 ft 8 5 m at 30 inHg 762 mmHg The maximum suction head is the vertical height at which a sample can be taken Pump Tube Life Tube life may vary depending on the abrasiveness of the sample liquid Recommended maximum 1 000 000 pump counts This equates to 912 ...

Page 28: ... mA CDMA cellular telephone modem standby current 35 mA current while communicating 150 mA GSM cellular telephone modem standby current 62 mA current while communicating 230 mA a Current ratings when input power is 12 5 VDC and ambient temperature is 77 F 25 C b Average current when programmed to take readings at 15 minute intervals c Rating for three outputs Sampler Only Operating Current Approxi...

Page 29: ...g level 0 0001 m storage resolution Bubbler 2 bytes representing level 0 0001 m storage resolution Area Velocity 2 bytes representing level 0 0001 m storage resolution 2 bytes representing velocity0 001 ft s storage resolution 4 20 mA 2 bytes representing percent0 1 storage resolution 2 bytes representing level 4 bytes representing flow 1 byte 8 bits SDI 12 Sonde Readings Parameter Range Resolutio...

Page 30: ...0 degrees Rectangular with and without end contractions Cipoletti Flumes Parshall 1 2 3 6 9 1 5 2 3 4 5 6 8 10 12 Palmer Bowlus 4 6 8 12 15 18 21 24 27 30 48 Trapezoidal LG 60 V 2 45 WSC 12 45 SRCRC H 0 5 0 75 1 1 5 2 2 5 3 4 5 Manning Formula Round U channel Rectangular Trapezoidal Area Velocity Round U channel Rectangular Trapezoidal Data Points 50 level flow rate points 50 level area points Flo...

Page 31: ... mode 12 VDC 6 amperes from external battery AC mode 90 264 Volt 47 63 Hz 2 amperes Operational Temperature 32 to 122 F 0 to 50 C Bottle Configurations 14 wedge polypropylene bottles 950 mL 1 quart 4 polypropylene bottles 5 0 liter 1 3 gallon 4 glass bottles 1 8 liter 5 gallon 1 round polyethylene bottle 9 4 liters 2 5 gallon 1 round glass bottle 9 4 liters 2 5 gallon 1 square polypropylene bottle...

Page 32: ...in I 5V Programmable I O3 pin Interrogator 6 Pin Female Sealed A Switched 12 volts DC B Ground C Connection Sense D Transmit E Receive F NC Flow Meter 6 Pin Male Sealed A 12 volts DC B Ground C Flow Pulses In D Bottle Number Out Dual Sampler Out E Event Mark Out F Inhibit In Module in module bay 8 Pin Male Sealed 1 NC 2 12 volts DC 3 Ground 4 Module Sense 5 Receive 6 Transmit 7 Clock 8 Control 12 ...

Page 33: ...pler Section 1 Introduction 1 17 12 V BACKUP POWER 2 PIN MALE SEALED A Ground B 12 volts DC 5 PIN MALE SEALED A 12 volts DC B unassigned C Enable signal D Ground E unassigned Table 1 4 Connector Specifications Continued REFRIGERATION ...

Page 34: ...BLZZRD Portable Refrigerated Sampler Section 1 Introduction 1 18 ...

Page 35: ...en the protective cap on this Flowmeter connector or attach a mating cable to it This connector must be sealed to prevent moisture from entering the case The controller case ventilation prevents differences between internal case and atmospheric pressures Extreme differences in pressures will affect the keypad label If bubbles form under the keypad label momentarily open this vent by removing the c...

Page 36: ...location to install the sampler or suction line observe standard safety precautions Refer to Appendix C for a discussion of safety considerations The following points should also be considered Power The only means to totally remove power from the sampler is by disconnecting the mains line cord from the power outlet or the cable to the battery Position the battery or connect to the mains outlet in ...

Page 37: ...n sulfide gas marine atmospheres or salt spray will damage the refrigeration system Avoid submersion Although the controller will resist damage rated NEMA 4x 6 the refrigerator system and bottle compartment cannot prevent the liquid from entering Liquid entering the refrigerated system will damage the cooling system liquid entering the bottle compartment will contaminate the collected samples Acce...

Page 38: ... damage to the sampler Check the condition of the pump tube regularly and if the tube shows signs of fatigue or wear replace it immediately A properly maintained sampler will pro vide years of reliable service that is expected of a Teledyne ISCO Sampler Section 9 4 of this manual describes the pump tube removal and replacement steps Afterwards be sure to reset the pump counter section 9 3 2 2 4 1 ...

Page 39: ... kit keep these guidelines in mind Install all bottles to ensure that they remain in place Align the bottles correctly If the bottles are misaligned the sampler may miss the bottle mouth or the sampler will deposit samples in the wrong bottle Each time you change bottle configurations check the program settings for the number of bottles and bottle volume 2 5 1 Installing the 14 Bottle Kit To insta...

Page 40: ...o the carrier Note that the bottles fit between the handles of the carrier The handles keep the bottles positioned correctly in the carrier 6 Place the carrier into the refrigerated compartment The carrier must be aligned with the alignment post on the plate See Figure 2 4 Once it is aligned it should seat fully against the adapter plate 7 Close and latch the refrigerated compartment 1 2 3 4 5 1 B...

Page 41: ...r Section 2 Installation Preparation 2 7 Figure 2 3 14 Bottle adapter plate installed Figure 2 4 Carrier alignment Figure 2 5 14 Bottle configuration completed 1 1 1 Alignment Posts 1 2 3 1 Bottle Carrier 2 Adapter Plate 3 Alignment Post ...

Page 42: ...ibutor arm Then place the end of the discharge tube through the opening in the end of the distributor arm The end of the tube should protrude below the arm 1 16 to 1 8 inch see Figure 2 2 If too much tube protrudes from the arm it can catch on bottles and jam the distributor arm 4 Install the 4 bottle adapter plate in the refrigerated com partment The bottom of the compartment has two posts that e...

Page 43: ...ure 2 7 4 Bottle adapter plate installed Figure 2 8 4 Bottle configurations completed 1 Alignment Posts Left Four 5 L Plastic Bottle Adapter Plate Right Four 1 8 L Glass Bottle Adapter Plate 1 1 Left Four 5 L Plastic Bottle Configuration Right Four 1 8 L Glass Bottle Configuration ...

Page 44: ...ne 68 2970 009 Refer to Figures 2 9 and 2 10 Both 2 5 gallon composite bottle kits use the same adapter plate Figure 2 9 Composite 2 5 gallon glass kit Figure 2 10 Composite 2 5 gallon Nalgene polyethylene kit 1 2 3 4 1 Glass Bottle 2 Composite Tube Guide 3 Adapter Plate 4 Discharge Tube 2 5 Bottle Lid not shown 1 2 4 3 1 Nalgene Bottle 2 Composite Tube Guide 3 Adapter Plate 4 Discharge Tube 2 5 B...

Page 45: ... into the side opening of the composite tube guide The end of the tube should be posi tioned just below the bottom opening of the composite tube guide 4 Install the composite bottle adapter plate in the refriger ated compartment The bottom of the compartment has two posts to ensure that the adapter plate is oriented cor rectly See Figure 2 12 5 Place the bottle into the refrigerated compartment Th...

Page 46: ...ZZRD Portable Refrigerated Sampler Section 2 Installation Preparation 2 12 Figure 2 12 Composite bottle adapter plate installed Figure 2 13 Composite 2 5 gallon bottle kits completed 1 1 1 Alignment Posts ...

Page 47: ...and 5 gallon composite bottle into the refrigerated compartment The bottle opening should be oriented toward the left side of the compartment Note The 5 gallon configuration does not use a composite tube guide 3 While closing the refrigerated compartment place the end of the discharge tube into the bottle opening 4 Close and latch the refrigerated compartment Figure 2 14 Composite 5 gallon bottle ...

Page 48: ...ch to DC power 2 6 1 Installing the Battery Box To fasten the battery to the tray use the battery box strap that is supplied with the 948 battery box to retain the battery in place Pre drilled holes are in the battery tray to allow for the fastening of the battery box strap Alternatively use an elastic bungee and place the hooks into the battery tray holes Figure 2 16Mounting holes for battery box...

Page 49: ...eep cycle marine battery with heavy duty clips These cables provide over current protection through the use of an in line 8A Slo Blo fuse CAUTION Only use Teledyne ISCO battery cables to connect the sampler to a DC power source The cable length and fusing protect you and the equipment from over current conditions and the risk of fire CAUTION Never use a DC extension cable without first consulting ...

Page 50: ...desired length but use the shortest length feasible for the installation Cut the line in 1 foot increments For instance 4 feet not 31 2 If using metric units of measure cut the line in increments of 0 1 meter Do not include the length of the strainer in the measurement The vinyl line contains a very low parts per million level of phenols If phenol content affects your sample analysis use the PTFE ...

Page 51: ...aded con nector into the suction line If attaching the strainer to a PTFE suction line heat the end of the PTFE suction line before threading the strainer into the line Warming the line will make the line more pliable to avoid damage 2 8 1 Alternative to Strainers When sampling from high velocity streams with heavy sus pended solids some field investigations suggest that more repre sentative sampl...

Page 52: ...tions 2 8 3 Pressurized Lines The sampler can obtain samples from pressurized lines as long as the line pressure remains below 15 PSI pounds per square inch Pressures greater than 15 PSI may prevent the sampler from purging the suction line moreover extreme pressures can force liquid past the pump even when the pump is not running For more consistent sample volumes from lines under constant pressu...

Page 53: ...dge Before installing one of these modules remove the desiccant cartridge and the two retaining clips This is necessary for the module to fit under the controller cover After installing the module insert the desiccant cartridge into the sampler s two retaining clips located behind the controller Refer to the Module s manual for detailed installation instruc tions Generally to install a module 1 If...

Page 54: ...tion 5 16 For complete SDI 12 sonde setup instructions refer to Section 7 Interrogator Connector This connector is the primary port for data collection and external sampler control Note Connecting any device to this connector disables the second ary communication port the factory installed dialout modem If you intend to use the optional dialout modem for data collec tion or instrument control ensu...

Page 55: ...not certified for use in hazardous locations as defined by the National Electrical Code Contact your Teledyne ISCO representative for your equipment s status Never operate equipment in hazardous locations that has not been certified as safe for such use Refer to Appendix C in the back of this manual for more safety information 2 10 Programming the BLZZRD Due to the many features supported by the B...

Page 56: ...otected Functions The user selected password defaults to 6712 and may be changed in the Software Options that are available in the Extended Program mode While the password may only be changed in the Extended Program mode the program lock works for both Standard and Extended programs Should you forget your password a key code can be obtained from Teledyne ISCO s Customer Support department for a no...

Page 57: ...eplace it with a fully charged battery The sampler has a fuel gauge that gives an indication of power usage The controller keeps track of how much power has been consumed since the last time it lost power The current power consumption as well as the previous power consumption can be accessed by pressing the STOP key while in the main menu Keep in mind that the sampler has no idea as to how much ch...

Page 58: ...BLZZRD Portable Refrigerated Sampler Section 2 Installation Preparation 2 24 ...

Page 59: ...at are common to both programming modes such as using the menus and setting the clock CAUTION This sampler has a high performance pump As with all such pumps it relies upon liquid to cool working components If the sampler is programmed to pump in the absence of liquid in excess of 5 minutes excessive heat buildup may damage the paddles rollers and housing Ensure that the liquid inlet is completely...

Page 60: ...t the Keypad KEY NAME FUNCTION Standby Turns sampler on or off Stop Stops the pump or distributor Pauses a running sampling program In programming screens returns to a previous screen Enter Accepts a menu choice or number entry and goes to next screen Help In programming screens displays a brief help message Down Right Arrow Selects the menu option right or below the current choice Up Left Arrow S...

Page 61: ...ear on your unit When the language choice you want is blinking press Enter The key always accepts the blinking option 3 2 Using Menus and Entering Numbers A menu is a list of options The main menu has four options The options identify the operations that can be performed from the menu For example to run the sampling program select RUN to modify the program select PROGRAM 3 2 1 Selecting Menu Optio...

Page 62: ... to go to the next screen Press Stop to return to the main menu Because the forward arrow is always blinking when the screen appears you can simply press to go to the next quick view screen making it easy to page through the screens Example Entering Numbers To enter a number Press the number keys to type the number Then press As soon as you press the sampler saves your number and moves to the next...

Page 63: ...reen containing the sample volume setting 2 The down arrow will be blinking Continue paging through the screens by pressing the key The Sample Volume screen will appear after several pages 3 Press either Arrow key until the 200 ml SAMPLES line blinks and press the key The sampler displays the normal view screen so you can view and modify the setting 4 After modifying the setting press the key The ...

Page 64: ...w setting For example to enter 2 00 p m 2 00 p m is 14 00 on a 24 hour clock type 14 Press Next type 0 zero for the minutes and press To enter a date such as January 22 2004 type 22 01 04 3 5 ClockStartTimeMenu Screens There are three start time settings you can use in a sampling program This section discusses one of these settings the clock start time A clock start time lets you set up a sampling...

Page 65: ... has two text entry menus one for site description and another for an extended program name They work in the same way A site description is commonly a number address or other short note that helps identify the monitoring site Extended programming has a similar screen that lets you enter program names for stored programs The example Entering a Site Description shows how to change the site descripti...

Page 66: ... the original description or program name unchanged Example Entering a Site Description The steps below show how to change the site description from FACTORY01 to SITE 29 These steps are similar to changing an Extended Program Name 1 From the Main Menu select PROGRAM 2 Select YES to change the Site Description 3 The far left character is blinking Use the arrow keys to move the cursor to the desired...

Page 67: ...d forth between screens by pressing the arrow keys 4 Press Stop at any time to return to the programming screen Most help notes refer to a related section in this manual Screens that require extensive explanation have notes that contain only a reference to the manual Sections mentioned in the notes appear in the index and the table of contents Example Viewing Help Notes This example shows how to v...

Page 68: ...e sampler to overfill the bottle 3 7 4 Operation Messages An operation message reports the sampler s status as it runs a program An example is shown below 3 8 Menu Trees Selecting a menu option will take you to a number entry screen or another menu screen The screens are organized in a branching structure that resembles a tree Refer to Figure 3 1 for a simple chart of the menu tree for standard pr...

Page 69: ...L A R M I N T E R N A L B A T T E R Y D I A G N O S T I C S V I E W D A T A P R I N T D A T A S Y S T E M I D s C O N F I G U R E R E P O R T S M A I N T E N A N C E M A N U A L F U N C T I O N S P R O G R A M M I N G S T Y L E P R O G R A M M I N G S T Y L E N O R M A L Q U I C K V I E W C H A N G E See Section 6 Running Programs See the screen charts in Appendix A See Section 6 7 Reports See Fig...

Page 70: ... over That is the sampler will replace the oldest stored event with the newest event When using an SDI 12 sonde the sampler logs additional entries at each sample event Module sonde refrigerator temperature and rain gauge readings are stored in memory partitions The storage capacity depends on the reading types the number of bytes per reading and the data storage interval The sampler clears the sa...

Page 71: ...wo minutes extends the storage capacity to 201 days The sampler does not have to be running a program to store readings When the sampler is on it stores readings at the selected data storage interval Once the partition is filled the readings roll over When the sampler is running a program the partitions become triggered by the programmed start time or the first enable time whichever is later A tri...

Page 72: ...BLZZRD Portable Refrigerated Sampler Section 3 Programming Introduction 3 14 ...

Page 73: ...andard programming mode The start up screen tells you which programming mode the sampler is currently using You can switch between programming modes by entering a numerical command at the main menu Type 6712 1 to enter standard programming mode Type 6712 2 to enter extended programming mode 4 2 Selecting Language and Units of Length You have the option of displaying screens in English or Spanish o...

Page 74: ...her arrow key until the correct number blinks Press For this example select 14 6 Type the volume for the bottles in your kit For this example 950 is correct so simply press 7 Type the length of the suction line then press If you change the length the sampler will display a message PLEASE WAIT GENERATING PUMP TABLES 8 Because this example requires samples every 15 minutes select TIME PACED by press...

Page 75: ... press 13 For this example select DELAYED START by pressing an arrow until the option blinks Then press 14 Type the delay period you want between the time you run the program and the time the sampler takes the first sample Then press 15 Run the program immediately by selecting YES Select NO if you want to run the program later by selecting RUN from the main menu Press after making your choice In t...

Page 76: ...e blinking Press 3 Select the desired temperature units 4 Select 14 Press 5 Type 950 and press 6 Type 7 then press 7 Select FLOW PACED then press 8 Enter the number of pulses between sample events Press 9 Select BOTTLES SAMPLE 10 Enter 2 11 For this example select NO to disable the Run Continuously option Selecting YES allows the program to run indefinitely by repeating the sample distribution and...

Page 77: ...ed sampling requires a flow meter flow logger or a 700 Series flow module A flow meter or flow logger paces a sampler by sending an electronic signal to the sampler after measuring a specified volume of liquid Because each pulse represents a volume interval flow pacing rates are proportional to the volume of water flowing through the channel Standard Program Flow Paced Sampling Two Bottles Per Sam...

Page 78: ...g The flow meter or flow logger sends pulses at two different time intervals when trigger pacing a sampler Trigger pacing for example lets you sample less frequently when the trigger condition a con dition based on level flow or rainfall remains below a set point The intervals are determined by the flow instrument s program settings for trigger pacing When conditions change and readings pass the s...

Page 79: ... B U T I O N 2 B O T T L E S S A M P L E STANDARD PROGRAMMING Bottles per Sample 1 1 2 2 3 3 S E Q U E N T I A L B O T T L E S S A M P L E S A M P L E S B O T T L E 2 S A M P L E S P E R B O T T L E 1 5 0 D I S T R I B U T I O N 2 S A M P L E S B O T T L E STANDARD PROGRAMMING Samples per Bottle 1 3 5 7 2 4 6 8 R U N C O N T I N U O U S L Y Y E S N O Each bottle receives one sample from one sample...

Page 80: ... configurations samples per bottle distribution is known as composite sampling A composite sample represents an average of the flow stream s characteristics during the sam pling period 4 5 5 Continuous Sampling Sample programs can be run indefinitely by selecting YES at the RUN CONTINUOUSLY screen Continuous sampling resets the distribution when the distribution sequence is com plete That is when ...

Page 81: ...e The sampler must be enabled and the pacing selected must call for a sample at start In any event the sampler checks the pacing set tings and begins the pacing interval countdown Example Start Times Delayed Starts count down before running a sampling program To enable a Delayed Start 1 Select DELAYED START 2 If the program is time paced enter the number of minutes to delay the first sample If the...

Page 82: ...he sampler s responses Figure 4 2 Start Time Diagram 4 7 Programming for 700 Series Modules The bay on the controller s side accepts any of Teledyne ISCO s 700 Series Modules These modules are optional accessories that are not required for operation of the sampler However the modules offer an economical way to combine flow rate or parameter monitoring with sampling The programming of these modules...

Page 83: ...rogramming the sampler and module Refer to the example Standard Program Flow Paced Sampling Two Bottles Per Sample which does not have a module attached at the beginning of this section to compare the sequence of screen displays CAUTION Most Teledyne ISCO accessories and interfacing equipment are not certified for hazardous locations as defined by the National Electrical Code Check applicable guid...

Page 84: ... Select the desired units for the flow volume For this example choose Mgal Press 9 Select YES to program the module 10 If you choose FLOWMETER proceed to step 11 If you choose LEVEL ONLY proceed to step 14 11 When you select Flowmeter you will have several options for performing the flow conversion For this example choose Weir Press Continued MODULE INSERTED DOWNLOAD DATA NOW OR LOSE ALL DATA DONE...

Page 85: ...n 19a If the module is being used for level only you will see screen 19b 19 If the module is being used as a flowmeter choose the amount of flow between sample events If level only enter the number of pulses between sample events Flow pulses must be provided by an external flow meter 20 Select BOTTLES SAMPLE 21 Enter 2 22 Select NO 23 Enter 250 24 Select NO DELAY TO START 25 Enter 0 This setting w...

Page 86: ...tion FLOWMETER or LEVEL AND VELOCITY For this example choose FLOWMETER 12 When you select Flowmeter you will have two options for performing the flow conversion AREA VELOCITY or LEVEL TO FLOW RATE For this example choose AREA VELOCITY 13 For this example choose Channel Shape 14 For the channel shape select Round Pipe 15 Enter the diameter of the round pipe This example shows a two foot diameter pi...

Page 87: ...u sure Yes No Choose NO to re enter select YES if correct 17 Enter the amount of zero level offset Refer to the 750 Area Velocity manual for more information on zero level offsets 18 If you are using a standard size sensor select the maximum depth for the velocity measurement This screen does not appear if using a low profile sensor which is automatically set to 1 inch 19 From this point forward t...

Page 88: ...BLZZRD Portable Refrigerated Sampler Section 4 Standard Programming 4 16 ...

Page 89: ... 4 5 1 Switching Between Standard and Extended Modes As a factory default the first time you turn the sampler on it will begin in standard programming mode The start up screen tells you which programming mode the sampler is currently using You can switch between programming modes by entering a numerical command at the main menu Type 6712 1 to enter standard programming mode Type 6712 2 to enter ex...

Page 90: ...lso save your 700 Series Module program settings and any sampler enable conditions using rain gauge data or SDI 12 sonde parameters These program settings remain as long as you do not change the type of module or change the rain gauge or SDI 12 Hardware Setup 5 3 1 Selecting a Stored Program When the sampler is in extended programming mode it always has one current and three noncurrent programs ea...

Page 91: ... type 6712 2 at the main menu shown in step 2 2 Select PROGRAM from the main menu 3 Select PROGRAM NAME EXTENDED 1 from the quick view menu You can go through QUICK VIEW screens quickly by pressing at each screen You can also change settings in quick view screens see section 3 3 Quick View Screens 4 Select SELECT NEW PROGRAM to change the program 5 Select the name of the extended program you want ...

Page 92: ...splayed depending on the type of module and the selections you make while programming Two modules that are commonly used are the 730 Bubbler Module which monitors a flow stream s level and flow rate and the 750 Area Velocity Module for monitoring a flow stream s level velocity and flow rate Section 4 contains two examples of programming for modules Module programming is the same in Extended as it ...

Page 93: ...le enable controls that is when the rain gauge detects 0 15 inches of rainfall in 30 minutes and the module detects a flow stream level of more than 0 25 feet Because both program parts use the same enable settings the sampler will enable both parts at the same time A list of settings for the program follow Program type Extended Program name STORM Site description SITE 54 Units Length ft Flow Rate...

Page 94: ...extended programming follow the example Example Storm Water Runoff Sampling Program 1 This screen will appear briefly Note that controller is in the EXTENDED PROGRAMMING mode 2 Select PROGRAM from the main menu 3 Select YES Note If the quick view screen shows up press STOP then select OTHER FUNCTIONS SOFTWARE OPTIONS QUICKVIEW CHANGE NORMAL Then press STOP and go back to step 2 4 Select CHANGE PRO...

Page 95: ... the stored data will be lost Download stored data if desired and select DONE 20 Select 14 21 Enter 950 Continued SITE SITE 54 ABCDEFGHIJKLMNOPQRST UVWXYZ 0123456789 BACK UP DONE SELECT UNITS FOR LENGTH ft m SELECT UNITS FOR LENGTH F C SELECT UNITS FOR FLOW RATE cfs gps gpm Mgd lps m3s m3h m3d SELECT UNITS FOR FLOW VOLUME cf gal Mgal m3 lit PROGRAM MODULE YES NO MODE OF OPERATION FLOWMETER LEVEL O...

Page 96: ...RAIN Note All valid enable options will appear here so your screen may differ 37 Select AND Continued SUCTION LINE LENGTH IS 5 ft 3 99 AUTO SUCTION HEAD ENTER HEAD 0 RINSE CYCLES 0 3 RETRY UP TO 0 TIMES WHEN SAMPLING 0 3 ONE PART PROGRAM TWO PART PROGRAM 14 BOTTLES AVAILABLE ASSIGN BOTTLES 1 THRU 6 TO PART A 1 13 UNIFORM TIME PACED FLOW PACED EVENT PACED NONUNIFORM TIME TIME BETWEEN SAMPLE EVENTS ...

Page 97: ...utes the delay to start of sampling specified in Step 68 after the enable for Part B 52 Enter 1 Continued ENABLE RAIN AND LEVEL FLOW FR TEMP DONE RAIN SET POINT 0 15 INCHES PER __HOURS __MINUTES ENABLED WHEN ABOVE SET POINT BELOW SET POINT RESET RAIN HISTORY YES NO LEVEL CONDITION SET POINT RANGE RATE OF CHANGE LEVEL SET POINT 0 25 ft 0 001 30 000 ENABLED WHEN ABOVE SET POINT BELOW SET POINT ONCE ...

Page 98: ...Enter 20 Continued SWITCH BOTTLES ON NUMBER OF SAMPLES TIME SWITCH BOTTLES EVERY 1 SAMPLES 1 47 RUN CONTINUOUSLY YES NO SAMPLE VOLUME 500 ml 10 950 ENABLE RAIN LEVEL FLOW FR TEMP A DONE NONE ENABLE RAIN AND OR DONE ENABLE RAIN AND LEVEL FLOW A DONE DONE RAIN SET POINT 0 15 INCHES PER __HOURS __MINUTES ENABLED WHEN ABOVE SET POINT BELOW SET POINT RESET RAIN HISTORY YES NO LEVEL CONDITION SET POINT ...

Page 99: ...le the sampler draws liquid through the line to the liquid detector As soon as it detects liquid the sampler reverses the pump to purge the line You can program the controller to perform as many as three rinse cycles before each sampling cycle Selecting a Stored Extended Program continued 69 Select DONE The screen will display End of Part B 70 Select DELAYED START 71 Select NO Option for samplers ...

Page 100: ...n be defined for two part pro grams up to fourteen event conditions can be defined for each part An event occurs when any of the programmed event condi tions becomes true By default event paced sampling uses a sequential distribution scheme meaning that each time an event occurs the sampler takes one sample and places it in one bottle However any distribution setting can be selected with event pac...

Page 101: ... for the upper range 13 Choose INSIDE RANGE 14 To stop entering events select DONE Proceed with the rest of the programming steps UNIFORM TIME PACED FLOW PACED EVENT PACED NONUNIFORM TIME EVENT01 RAIN LEVEL FLOW FR TEMP pH5 SP_CO DO5 LEVEL CONDITION SET POINT RANGE RATE OF CHANGE LEVEL SET POINT 1 000 ft 0 001 30 00 ENABLED WHEN ABOVE SET POINT BELOW SET POINT EVENT02 RAIN LEVEL FLOW FR TEMP pH5 S...

Page 102: ...g Nonuniform clock time pacing Nonuniform interval pacing Random interval pacing 5 8 3 Nonuniform Clock Time Pacing For clock time pacing enter specific times for each sample event see screen 2 in the example You can enter as many as 99 clock times but the interval between times cannot be greater than 24 hours Remember to use a 24 hour clock when entering times The sampler takes a sample at the st...

Page 103: ...om the distribution settings Nonuniform Time Interval Pacing allows you specify the time interval between each sample 1 Select INTERVALS IN MINUTES 2 The controller briefly displays then displays the screen to enter as many as 99 nonuniform time intervals Enter the number of samples to be taken at each interval then specify the interval in minutes NONUNIFORM TIME CLOCK TIMES INTERVALS IN MINUTES R...

Page 104: ...before moving to the next bottle Use samples per bottle distribution to collect a series of small composite samples 5 9 4 Composite For single bottle configurations samples per bottle distribution is known as composite sampling A composite sample represents an average of the flow stream s characteristics during the sam pling period 5 9 5 Multiple Bottle Compositing Multiple bottle compositing is a...

Page 105: ...us Sampling Sample programs can be run indefinitely by selecting YES at the RUN CONTINUOUSLY screen Continuous sampling resets the distribution when the distribution sequence is com plete That is when the last bottle of the set is reached the next bottle is the first bottle of the set All pacing modes except RANDOM are supported When running a program continuously the bottles must be ser viced at ...

Page 106: ...O T T L E S E V E R Y 1 S A M P L E S 1 5 0 YES NO RUN CONTINUOUSLY Each bottle receives one sample from one sample event EXTENDED PROGRAMMING D I S T R I B U T I O N 2 B O T T L E S S A M P L E 2 B O T T L E S P E R S A M P L E S E V E N T 1 1 4 S W I T C H B O T T L E S ON N U M B E R O F S A M P L E S T I M E S W I T C H S E T S E V E R Y 1 S A M P L E S 1 5 0 YES NO RUN CONTINUOUSLY Multiple b...

Page 107: ... on an external flow measurement device attached to the Flow Meter connector select FLOW PULSES To base the volumes on the 700 Series Flow Module readings select FLOW MODULE VOLUME Next the sampler asks for the SAMPLE VOLUME to collect at each time interval This volume is expressed as 10 ml for every X number of pulses or if using a flow module for every volume unit Refer to section 5 10 1 Calcula...

Page 108: ...presentative of the flow stream At the lowest expected flow the SAMPLE VOLUME value should be set to ensure that enough sample is collected If too little sample liquid is collected there may not be enough for analysis although it is a representative composite sample For example consider an NPDES sampling program used to collect a flow proportional composite sample for BOD TSS and NH3 N analysis Th...

Page 109: ...on Because the SAMPLE VOLUME entry must be in whole numbers 16 7 is rounded down to 16 It is important to note that this number is inversely proportional to the amount of volume collected Rounding up would result in less volume collected The lowest possible SAMPLE VOLUME value is dependent upon the highest expected flow and the available bottle volume To cal culate the number of pulses or volume a...

Page 110: ...sures the stream s flow volumes The equation for the high end of the range is Using the same scenario the calculations would be The equation for the low end of the range is The scenario would be applied as Therefore a value should be selected between 709 and 1667 gallons for the given scenario Greatest Total Flow Vol Flow Pulse Vol Total Bottle Vol 10 ml per unit 10 ml for every __ pulses 666 667 ...

Page 111: ... an SDI 12 sonde at address 5 with four parameters Refer to Figure A 12 in Appendix A for details on programming sampler enables When the sampler detects that it is disabled at start time it sus pends the program until it is enabled Once enabled the sampler collects a sample only when programmed to take a sample at the enable time The diagram in Figure 5 4 shows the sampler s response when enabled...

Page 112: ...ET POINT 9 Enter 1 foot 10 Select ABOVE SET POINT 11 Your enable conditions will be summarized on the display If correct press to continue If necessary make changes to the enable conditions ENABLE NONE PROGRAMMED ENABLE RAIN LEVEL FLOW FR_TEMP pH5 SP_CO5 DO5 NONE ENABLE RAIN AND OR DONE ENABLE RAIN AND LEVEL FLOW FR TEMP pH5 SP_CO5 DO5 DONE RAIN SET POINT 5 INCHES PER __HOURS __MINUTES ENABLED WHE...

Page 113: ... enabled the sampler takes samples until it reaches the end of its program Note The STAY ENABLED feature works only when the sampler becomes enabled following program start If you need this fea ture with enable prior to program start contact Teledyne ISCO for details Repeatable Enable After enabling the sampler the enabling instrument or instruments continues to monitor the conditions and disables...

Page 114: ...mpler will wait for up to one year for the dry period to occur Resetting the Sample Interval at Enable You may want the pacing countdown to start over each time the sampler becomes enabled To accomplish this select YES when asked if you want to reset the sample interval at enable The next sample event will occur at the end of a full interval Furthermore if the sampler is enabled several times duri...

Page 115: ...sampling until the next resume time Pause and resumes repeat weekly You can enter up to nine pairs of pause and resume times The pause and resume times shown as examples below suspend the program between 5 00 PM Monday and 8 00 AM on Tuesday and between 5 00 PM Tuesday and 8 00 AM on Wednesday If the start time is Monday at 8 00 AM the sampler takes samples between 8 00 AM and 5 00 PM Monday betwe...

Page 116: ...s of the programmed enables You cannot change the type of enables used or the way they are combined If rain is an enabling condition you also have the option of resetting the rain history Adjust Suction Head If you have chosen to manually enter the suction head and are in the manual paused state the scrolling menu will have an option to adjust the suction head If you make changes to the suction he...

Page 117: ...Normally there are two considerations when selecting the Mea surement Interval power consumption and duration of the enabling event This setting for battery powered samplers will require some degree of balance between the need to conserve power yet detect the enabling event before it has passed 5 15 4 Dual Sampler Mode Dual Sampler Mode lets you operate two samplers as a pair The primary sampler d...

Page 118: ...tware setting after you select a different Extended Program This feature relies on a post purge after each sample When using the Bottle Full Detect make sure that post sample purges greater than 100 are used Bottle Full Detect is always disabled in the Standard Pro gramming mode 5 15 6 Display Backlighting The display has a backlight that is adjustable so that the back light is always off timed or...

Page 119: ...e 5 5 shows event mark timing diagrams Bottle Number Concurrent with the event mark the sampler also sends a bottle number signal The bottle number signal is a series of pulses The number of pulses sent represents the bottle currently accepting a sample see Figure 5 5 5 15 8 Pre sample and Post sample Purge Counts The pre sample purge is when the pump runs in reverse clearing the strainer of any d...

Page 120: ... receive serial data by command 1 Select NO at the screen asking for periodic serial output 2 Connect your computer to the sampler controller s interro gator connector with a standard interrogator cable P N 60 2544 040 25 pin or 60 2544 044 9 pin 3 Configure your communications software for a baud rate no higher than 19 200 no parity 8 data bits and 1 stop bit 4 From the computer s keyboard type r...

Page 121: ...DE Description BLZZRD Sampler FL Flow cubic meters per second ID Unit specific identifier 10 digit unsigned long LE Level meters MO Model BLZZRD PE Percent percent of full scale PH pH none PR Previous day s rain tips2 RA Rain tips2 RTE Refrigerator Temperature degrees celsius SS Sampler enable state logical TE Temperature degrees celsius TI Current time days since 1 Jan 1900 VE Velocity meters per...

Page 122: ...external pump or other device Alarm Dialouts Only Select this option to power this port only when a programmed dialout condition is true This option is intended for applications that use the optional internal dialout modem with a cellular phone To control the operation of other external devices based on alarm conditions use I O Pin Pro gramming instead LE_ 4 Level meters MM_ 4 Ammonium nitrogen mi...

Page 123: ...he rain gauge readings and add RAIN as a sampler enable option Select the appropriate tip for the rain gauge The sampler will then create a rain memory partition and update the available sampler enable conditions Select NO to disable the readings and the sampler enable option When you select NO the sampler removes the rain memory par tition leaving the memory space available for other reading type...

Page 124: ...larms The alarm output can be set for any RUNNING PROGRAM or RUN ERROR state In addition the CONDITIONS output can report alarm conditions using the parameters the 6712 is con figured to measure You can set thresholds for one or two condi tions and the CONDITIONS output can then trigger alarms when the measured parameters exceed the threshold Alarm function requires an alarm interface module facto...

Page 125: ...ent is connected to your sampler you may see other parameters in addition to RAIN such as LEVEL FLOW VELOCITY etc Under CONDITIONS you can program limits for a condition to trigger the alarm when the measured parameter exceeds its limit You can use one or two conditions to trigger one alarm using AND or OR operators 5 When you select a parameter with the arrow keys and Enter you can then program t...

Page 126: ...t when an alarm condition is true Before proceeding the I O conditions must be defined Refer to section 5 16 4 These steps program a dialout alarm when the condition defined for I O1 is true 1 Select YES to enable the dialout alarm The controller asks the same question for IO2 and IO3 Select NO 2 Using the number entry keys enter as many as 3 telephone numbers to the contact list If the phone numb...

Page 127: ...to select and enter the extended char acters Each comma entered will cause the modem to pause for two seconds before proceeding with the remaining digits Phone numbers that are no longer needed can be removed Advance the cursor to the end of the telephone number and then press the Left Arrow key to erase the preceding character If the cursor is not at the end of the telephone number the Left Arrow...

Page 128: ...he readings are stored as FR TEMP readings at the data storage interval You can retrieve a history of temperature readings through the sampler s reports The temperature data can also be used as a sampler enable condition On the sampler this hardware feature is always enabled to operate the refriger ation system 5 16 8 Analog Output Teledyne ISCO offers three analog outputs as a factory installed o...

Page 129: ...g screens 6 2 Run Time Screens While running a sampling program the sampler displays a variety of messages that report the program s status For example if the sampler has not yet reached the programmed start time it displays the scheduled start time as well as the current time Once the program reaches the start time you can determine the time of the next sample the next bottle to receive a sample ...

Page 130: ...s for more than five minutes for a level or velocity reading that flow is calculated from P R O G R A M M I N G C O M P L E T E R U N T H I S P R O G R A M N O W Y E S N O E N T E R S T A R T B O T T L E _ _ 1 M A X P L E A S E W A I T T E S T I N G D I S T R I B U T O R S Y S T E M O P E R A T I O N E R R O R D I S T R I B U T O R J A M M E D P R O G R A M W I L L S T A R T A T 0 6 0 0 M O 1 7 O ...

Page 131: ...isplay by selecting VIEW REPORT from the main menu 6 3 How Does theSampler Work When the sampler takes a sample it draws liquid through the strainer and suction line to the pump tube refer to Figure 6 2 The liquid flows through the pump tube and past the liquid detector which senses the liquid From the detector the liquid follows the pump tube through the pump to the bulkhead fitting and then thro...

Page 132: ...tings for distribution The sampler can move the distributor arm clockwise and counterclockwise making a number of distribution methods possible In this manual the words sample event refer to a full sampling cycle for any distribution Other variations include extended program settings for suction line rinses and sampling retries A sampler running a program with line rinses completes the pre sample ...

Page 133: ... conserving power The measured temperature is reported as FR_TEMP on the run state displays and in the stored data 6 5 Interrupting a Running Program You can interrupt a sampling program by pressing the Stop key while the sampler is waiting for the next sample event Pressing Stop places the sampler into Manual Paused operation and records a manual pause in the sample event log In this state the sa...

Page 134: ...e sampler base Do this by removing the pump tubing from the bulkhead fitting After the sample has been pumped you will then be asked to enter the amount actually delivered Once you have entered the delivered volume and accepted the calibration a volume delivered event will be logged See Section 6 6 2 for more infor mation 6 5 7 Cal Adj Parameters When the sampler is configured for operation with a...

Page 135: ...r has been consumed since the last time it lost power If the con troller is powered by an external battery this fuel gauge can help you estimate the condition of the battery Keep in mind that the sampler has no idea as to how much charge was originally stored in the battery or even if a battery is being used and therefore has no idea as to the remaining capacity of its power source 6 6 Manual Func...

Page 136: ...for line rinses or retries If it is a two part program the sampler uses the settings for part A To take a grab sample while the sampler is running a program you must interrupt the program by pressing the Stop key Restart the program by running it again To take a grab sample 1 Disconnect the pump tube from the bulkhead fitting 2 Place the end of the tube over a collection container 3 Initiate the g...

Page 137: ...f the pump tube over the container and press 4 The controller will complete a sample collection cycle 5 The collection cycle is complete when the display returns the MANUAL FUNCTIONS screen Press the Stop key to return to the main or paused menu GRAB SAMPLE CALIBRATE VOLUME OPERATE PUMP MOVE DISTRIBUTOR SAMPLE VOLUME ___ ml 10 9990 GRAB SAMPLE PRESS WHEN READY PURGING SUCTION LINE PUMPING 200 ml P...

Page 138: ...e calibrating The sample source is above the sampler Sampling from pressurized lines 15 psi maximum The controller has been reinitialized Reinitializing the controller clears the calibration data Volumes cannot be calibrated while a program is running The program must be paused or stopped first For best results Calibrate after the sampler has be installed on site Use a graduated cylinder for volum...

Page 139: ...me Measure the actual volume delivered to the collection container and enter the amount here 6 If you enter a volume more than twice or less than half the programmed volume the message below is displayed Select YES when the volume delivered matches the volume displayed on this screen Select NO to re enter the volume delivered 7 The collection cycle is complete when the display returns the MANUAL F...

Page 140: ...ess the 3 key and then to run the pump forward CAUTION Prolonged exposure to noise while the pump is operating could result in hearing loss Tests indicate that the pump produces sound levels in excess of 85db at one meter If you are manu ally operating the pump for long durations always wear hear ing protection Example Operating the Pump Manually 1 Select OPERATE PUMP from the MANUAL FUNCTIONS scr...

Page 141: ...m example CAUTION Distributor drive gears may be permanently damaged if you attempt to rotate the distributor arm by hand or with any tool Only move the distributor arm through the Manual Functions Example Moving the Distributor Arm 1 Select MOVE DISTRIBUTOR from the MANUAL FUNCTIONS screen 2 Enter the bottle number and press 3 The distributor arm moves to the selected bottle position then display...

Page 142: ...Samplink to transfer the reports from the RTD to a file on the computer Note The RTD is not compatible with samplers running software ver sion 2 50 Refer to Section 1 4 5 for complete information Collect the reports except for summary reports with an IBM PC or compatible computer running Flowlink Collect the reports with and IBM PC or compatible computer running Samplink 6 7 2 Viewing the Data Bec...

Page 143: ...port options allow you to tailor the output to the Teledyne ISCO software used to collect the reports The Samplink Report option provides the Sampling Reports but disables the output of the detailed partition data Flowlink report provides the Sampling Reports and the detailed partition data Selecting the Flowlink report option enables the output of the detailed partition data so Flowlink can add t...

Page 144: ...ations in pump counts from event to event indicate fluctuating heads a relocated suction line or a temporarily clogged strainer 6 7 5 Combined Results Report At the time of each sample event the sampler records the readings from each connected module rain gauge or SDI 12 sonde The Combined Results Report presents the event time sample and bottle number and the respective readings from the device F...

Page 145: ... switch time Program Done The sampler finished the sampling program A B Done The sampler finished program part A or B Manual Pause Stop was pressed during the program to enter the Manual Paused state Program Stopped You selected Stop from the manual paused options terminating the program Grab Sample A grab sample was taken while in the Manual Paused state Pump Tube Replaced The pump tube alarm was...

Page 146: ...ample because the pump band was opened LF Not enough flow Not enough flow for the minimum sample volume entered for samples with volumes proportional to flow M Bottle Full The composite bottle is full ND No distributor arm There is no distributor arm or the stop is damaged NL No Liquid Detected The sampler was unable to detect liquid NM No More Liquid After the sampler detected liquid and while th...

Page 147: ...734 08 32 19 SEP 03 Hardware A1 Software 2 10 PROGRAM SETTINGS SITE DESCRIPTION FACTORY051 UNITS SELECTED TEMPERATURE F FLOW RATE cfs FLOW VOLUME Mgal BUBBLER MODULE WEIR 90 V NOTCH 14 950 ml BTLS 10 ft SUCTION LINE PACING TIME EVERY 2 HOURS 0 MINUTES DISTRIBUTION SEQUENTIAL 200 ml SAMPLES FIRST SAMPLE 08 00 MO TU WE TH FR ...

Page 148: ...t 07 53 TH 18 SEP 03 Nominal Sample Volume 200 ml COUNT TO SAMPLE BOTTLE TIME SOURCE ERROR LIQUID 08 00 PGM ENABLED 1 1 1 08 00 S 250 1 1 2 10 00 T 247 1 1 3 12 00 T 247 1 1 4 14 00 T 249 1 1 5 16 00 T 247 1 1 6 18 00 T 247 1 1 7 20 00 T 247 1 1 8 22 00 T 248 FR 19 SEP 03 1 1 9 00 00 T 237 1 1 10 02 00 T 236 1 1 11 04 00 T 237 1 1 12 06 00 T 241 1 1 13 08 00 T 241 08 06 PGM STOPPED 19 SEP SOURCE S...

Page 149: ...OW TOTAL LEVEL RATE FLOW SAMPLE BOTTLE TIME ft cfs Mgal 1 1 1 08 00 0 009 2 495 000000 000 1 1 2 10 00 0 012 2 576 000000 102 1 1 3 12 00 0 012 2 666 000000 262 1 1 4 14 00 0 012 3 162 000000 396 1 1 5 16 00 0 012 3 341 000000 556 1 1 6 18 00 0 012 3 713 000000 764 1 1 7 20 00 0 012 5 428 000000 928 1 1 8 22 00 0 012 6 864 000001 128 FR 19 SEP 03 1 1 9 00 00 0 012 3 339 000001 306 1 1 10 02 00 0 0...

Page 150: ... 200 ml FR TEMP SAMPLE BOTTLE TIME F 1 1 1 08 00 41 4 1 1 2 10 00 34 2 1 1 3 12 00 34 4 1 1 4 14 00 34 5 1 1 5 16 00 34 5 1 1 6 18 00 34 3 1 1 7 20 00 34 3 1 1 8 22 00 34 1 FR 19 SEP 03 1 1 9 00 00 34 1 1 1 10 02 00 34 2 1 1 11 04 00 34 3 1 1 12 06 00 34 4 1 1 13 08 00 34 2 SAMPLER ID 3687447734 08 32 19 SEP 03 Hardware A1 Software 2 10 COMBINED RESULTS SITE FACTORY051 Program Started at 07 53 TH ...

Page 151: ...0 NO DATA 01 00 02 00 NO DATA 02 00 03 00 NO DATA 03 00 04 00 NO DATA 04 00 05 00 NO DATA 05 00 06 00 NO DATA 06 00 07 00 NO DATA 07 00 08 00 NO DATA 08 00 09 00 2 519 cfs 09 00 10 00 2 447 cfs 10 00 11 00 2 591 cfs 11 00 12 00 2 552 cfs 12 00 13 00 2 721 cfs 13 00 14 00 2 978 cfs 14 00 15 00 3 096 cfs 15 00 16 00 3 223 cfs 16 00 17 00 3 402 cfs 17 00 18 00 3 668 cfs 18 00 19 00 3 811 cfs 19 00 20...

Page 152: ...hat data from different samplers is not combined in the same file The software version number changes when your sampler s software is updated The system will also list any software options that you have acti vated on your unit These specialized software options are available for purchase from Teledyne ISCO s Special Product Applications Department Example Viewing System IDs 1 Select View Report fr...

Page 153: ...l water up to 50 degrees C and 49 feet 15 meters depth Its ease of use design includes direct USB connection to your com puter real time data viewing and LED indicators for operating status Using SDI 12 protocol the AQ700 can connect to the fol lowing Teledyne ISCO instru ments 6712 or BLZZRD Waste water Sampler Signature Flow Meter 2105 Interface Module There are two versions of the AQ700 to choo...

Page 154: ...uipment could record the data values for one parameter as another s such as reporting a pH value as being temperature Some sondes such as the AQ700 have plug and play setup Known as ISCO Ready these sondes can tell the sampler what values they have their order and units of measure Other SDI 12 compatible sondes can still be used but require manual setup to identify proper data types for each data ...

Page 155: ...an at startup is turned off These settings are the factory defaults If you have made changes to the settings or have previously configured a sonde it may appear different than above 3 Use the arrow keys to select NO SDI 12 SONDE and press Enter The next display indicates that it is scanning for sondes Table 7 1 SDI 12 Sonde Parameters Parameter Units Range Resolution Bytes Temperature C 40 100 0 1...

Page 156: ...able for use in the sampling programs Section 5 describes the Extended Programming features Note that after a sonde has been found and configured the SDI 12 HARDWARE screen will no longer appear as it did in step 2 Instead it will report the available SDI 12 data similar to the screen below Selecting SDI 12 DATA will give you the following options NO SDI 12 SONDE select this option to remove the s...

Page 157: ... 1 Turn off the sampler controller and connect the SDI 12 sonde to the rain gauge connector 2 Turn on the sampler controller 3 From the EXTENDED PROGRAMMING mode s Main Menu select OTHER FUNCTIONS then HARDWARE Select NO SDI 12 SONDE 4 The controller searches sondes 5 When it finds a sonde it will briefly report the number of sondes found You can wait for the screen to advance or press 6 Select th...

Page 158: ...l appear as OTHER You can label the parameters so the BLZZRD controller will display the parameters and measurement units correctly To do so you must know what parameters are reported by the sonde This is typically documented by the manufacturer of your SDI 12 device or available by connecting to the device using a terminal communications program from a personal computer Refer to your SDI 12 devic...

Page 159: ... the main menu then choose Hardware Advance through the screens until you see one that displays SDI 12 DATA Select this screen to display the following Select SETUP KNOWN SONDES and then indicate the address of the sonde you want to setup A screen containing the sonde model and number of data values will appear briefly Then you will be asked to verify parameter by parameter what data you want to s...

Page 160: ... solution and verify the sonde s readings The validation process does not modify any readings like calibration does it merely alerts you to inaccurate ones No data will be stored nor will enable conditions be updated while calibration or validation is being performed 7 7 1 Calibration Screens To access the calibration screens from the main menu select Other Functions Hardware Advance to SDI 12 DAT...

Page 161: ... reading will appear at the bottom of the screen Select DONE to return to the Hardware option screens 7 8 Disconnecting Sondes When you have disconnected all sondes and have interrogated the sampler for all its data you can indicate that there is no SDI 12 sonde From the main menu select Other Functions Hardware Advance through the screens until you see one that displays SDI 12 DATA Select this sc...

Page 162: ...or much of its capabilities External program control ignores the sampler s program and current state Using external control you can turn on the sampler take a sample and request sampler status data The pump distributor and serial data output simply respond to the external commands To operate the sampler remotely on either level you must first establish RS 232 serial communication between the compu...

Page 163: ...mands can be entered from your computer These menu com mands are listed in Table 8 1 To access the menu commands type MENU at the prompt and press ENTER The screen will return a list of the available options Note that the current state of the controller will determine the available menu options Then type the desired command s confirming with ENTER Figure 8 2 Example list of remote menu commands ...

Page 164: ...turned A Remote Disable is the highest priority disable If the sampler is disabled with this command no condition other than menu command 6 ENABLE will allow the program to become enabled again 6 or ENABLE This command is used to re enable a disabled sampler ENABLE is valid while a program is running If successful a display of REMOTE ENABLE IS TRUE will be returned 7 or TAKE_SAMPLE This command ca...

Page 165: ...ing itself in the ON state and returning a status string to report the new status See Send Status for a description of this string The command to turn the sampler ON and return status is STS 2 CR Take a Sample This directs the sampler to take a sample The command specifies how much volume to take and which bottle to place the sample in The sampler responds by taking a sample and returning a status...

Page 166: ...red 5 PUMP JAMMED must be resolved before continu ing 6 DISTRIBUTOR JAMMED must be resolved before continuing 9 SAMPLER OFF 12 SAMPLE IN PROGRESS 20 INVALID COMMAND Possible causes may be identifier code is not supported bottle requested is not in current configuration sample volume requested is outside its range 10 9990 ml day Set_Time must be 5 digits and more recent than 1977 21 CHECKSUM MISMAT...

Page 167: ...y module rain gauge or SDI 12 device Refer to Table 8 2 to interpret the returned string Note This returns the same data as the programmable software option Serial Output as described in Section 5 17 6 of this manual The command to send data is DATA CR Table 8 2 Serial Data Codes Identifier Parameter Units B 1 Bottle number and time days since 1 Jan 1900 CR Current day s rain tips2 CS3 Checksum un...

Page 168: ...r liter MA_ 4 Ammonia nitrogen milligrams nitrogen per liter NI_ 4 Nitrate Nitrogen milligrams nitrogen per liter TB_ 4 Turbidity nephelometric turbidity units CL_ 4 Chloride milligrams per liter CP_ 4 Chlorophyll micrograms per liter FL_ 4 Flow rate cubic meters per second VO_ 4 Flow volume cubic meters xxD_ 4 5 Other data not listed above 1 The three most recent samples will have entries as B Th...

Page 169: ...format The checksum of 582 was calcu lated as shown below 8 1 3 Remote Control of Sampler Keypad To control the sampler keypad remotely type CONTROL at the prompt and press ENTER The sampler display appears on your computer monitor as you step through the programming screens While in this mode the computer keys will be redirected to simulate the sampler s keypad and the sampler keypad itself will ...

Page 170: ...s that a computer modem has called and initiates the modem connect sequence If this occurs you must hang up and try again 2 Enter the command For example to run program 1 press 3 If the command is valid for its current state of operation the sampler responds with command number please acknowledge If the command is not valid the sampler responds with the command number followed by three quick beeps...

Page 171: ...milar in functionality to a pin F low signal a signal typically used to enable or disable the sampler from a flow meter At run time the sam pler starts with the phone enable in the enabled state PHONE DISABLE is valid while a program is running If successful a reply of REMOTE ENABLE IS FALSE will be returned A Remote Disable is the highest priority disable If the sample is disabled with this comma...

Page 172: ...battery 9 2 Cleaning Guidelines Keeping the sampler clean and protected from harsh elements may extend the usable life of the sampler The following cleaning instructions can be performed as needed 9 2 1 Cleaning the BLZZRD Exterior The BLZZRD controller and refrigerator exterior may be cleaned with warm soapy water and a rag Never use acids or solvents to clean the sampler If there is excessive di...

Page 173: ...compartment Be sure to reinstall the plate in the same orientation Use the arrows to properly orient the bottom plate Figure 9 1 Align arrows when installing the bottom plate 9 2 3 Defrosting the Refrigerated Compartment Defrosting the refrigerator compartment may be necessary depending on the humidity of its operating environment If ice forms on the walls of the interior compartment never use sha...

Page 174: ...om previous sampling sites 9 3 Maintenance Screens When you select OTHER FUNCTIONS from the main menu and then select MAINTENANCE the following screen will be dis played Using these options you can set the sampler s time and date set the number of pump counts for the pump tube warning reset the internal battery count down timer and run a set of diagnostic tests for troubleshooting purposes 9 3 1 S...

Page 175: ...r briefly displays the current pump count information Line one lists the pump counts since the last reset Line two lists the current alarm setting The screen will advance automatically 5 To reset the counter to zero select YES Always reset the counter after replacing a pump tube Select NO when merely checking the current count 6 You can modify the pump tube alarm setting to tailor it to your needs...

Page 176: ...If it discharges completely the sampler will lose all program settings and all data stored in memory when the external power source is disconnected The battery requires replacement every five years The sampler keeps track of this five year replacement date which is calculated from the date the battery was installed Example Resetting the Internal Battery Countdown Timer 1 From the Main Menu select ...

Page 177: ...er it tests the pump and distributor and whether it reini tializes the unit to its original factory settings Example Self diagnostics 1 From the Main Menu select OTHER FUNCTIONS 2 Select MAINTENANCE 3 Select DIAGNOSTICS 4 The controller starts the diagnostics routine by first testing the RAM Random Access Memory If the controller passes a memory test it will advance to the next test Otherwise it w...

Page 178: ...quid while running the test and you can run the test with or without a tube in the pump Self diagnostics continued If you select YES the controller tests the forward and reverse pumping operation The ratio should fall within the 0 80 to 1 25 range Values outside this range indicate a pump problem contact Teledyne ISCO Technical Service 8 Select YES to test the distributor NO to skip this test If y...

Page 179: ...y recognized by the blue alignment collars Improper pump tubes include those made for Teledyne ISCO s other samplers 3700 Glacier etc and tubing from non Teledyne ISCO vendors Also note that the discharge tube is not the same as the pump tube You could experience several problems if you install the wrong pump tubing The sampler will not pump the liquid Pump jams Inaccurate sample volumes Faulty li...

Page 180: ...Close the pump strap and latch it 8 Reset the pump tube counter See the Resetting the Pump Tube Alarm example 9 Take a dry grab sample to test the tube installation WARNING The pump s safety lock prevents the sampler from running the pump when the pump band is open DO NOT tamper with the safety mechanism The pump is extremely powerful The pump rollers can injure you severely if the sampler activat...

Page 181: ...BLZZRD Portable Refrigerated Sampler Section 9 Maintenance 9 10 Figure 9 2 Removing and replacing the pump tube Figure 9 3 Placing the pump tube in the liquid detector ...

Page 182: ...es the arm to the distributor shaft If configured for 2 5 gallon composite bot tles remove the composite tube guide CAUTION Distributor drive gears may be permanently damaged if you attempt to rotate the distributor arm by hand or with any tool Only move the distributor arm through the Manual Functions 3 Remove the controller from refrigerator assembly by removing four screws See Figure 9 4 Figure...

Page 183: ...ture does accumulate the numbered areas on the indi cator will turn light pink or white starting with the area num bered 20 This indicates that the relative humidity inside the control box exceeds 20 As more moisture accumulates the areas numbered 30 and 40 will turn light pink or white indi cating relative humidities of 30 and 40 respectively If the 30 area of the humidity indicator turns light p...

Page 184: ...own Timer While replacing the battery it is likely that the stored readings will be lost and your program will revert to the default program Before proceeding retrieve the stored data and record your program settings Note Teledyne ISCO recommends you replace the battery every five years To ensure that the replacement meets Teledyne ISCO s specifications use only Teledyne ISCO replacement batteries...

Page 185: ...nector board 7 Turn the board over and locate the lithium battery on the component side Figure 9 8 Remove the lithium battery using care to prevent damage to the circuit board 8 De solder the cylindrical battery or the coin style battery Insert the replacement battery and solder it in place WARNING Danger of explosion if lithium battery is incorrectly replaced Replace with the same type ensuring c...

Page 186: ...connectors 10 Replace the grounding strap and attach with the 1 4 inch stop nut Reconnect the four connectors that connect the case bottom assembly 11 Ensure that all connections and hardware are secure Replace the desiccant and close the case 12 Mount the case on the refrigerator assembly 13 Apply power and turn the BLZZRD on 14 Reset the Battery Countdown See the example Resetting the Internal B...

Page 187: ... events have resulted in a POWER FAILED error message 9 8 1 Fatal Error Message The Fatal Error message will only show up if you are using a multiple bottle configuration and the distributor system fails Failing the test can indicate any of the following conditions The distributor arm is not attached to the sampler The distributor drive mechanism inside the controller is malfunctioning The control...

Page 188: ...I N G S T Y L E P R O G R A M M I N G S T Y L E N O R M A L Q U I C K V I E W C H A N G E P O W E R U S E D C U R R E N T _ _ a m p h r P R E V I O U S _ _ a m p h r S t o p K e y Note A Note A Note A Menu access can be controlled with the optional Program Lock See Section 5 See Section 6 Running Programs See Figure A 8 Manual Functions Screens See Flowcharts in Figures A 2 through A 21 See Figure...

Page 189: ...T A K E _ _ _ S A M P L E S 1 9 9 9 R U N C O N T I N U O U S LY Y E S N O S E Q U E N T I A L B O T T L E S S A M P L E S A M P L E S B O T T L E _ _ _ B O T T L E S P E R S A M P L E E V E N T 1 m a x _ _ _ S A M P L E S P E R B O T T L E 1 m a x 1 bottle No Yes Multi bottle B O T T L E V O L U M E I S _ _ _ _ _ l i t 0 3 1 0 0 0 i f 1 o r 4 b o t t l e s i f 1 4 b o t t l e s S E L E C T U N I ...

Page 190: ... N G C O M P L E T E R U N T H I S P R O G R A M N O W Y E S N O R U N P R O G R A M V I E W R E P O R T O T H E R F U N C T I O N S S T A R T E X T E N D E D 1 A F T E R A _ _ _ M I N U T E D E L A Y 1 9 9 9 S T A R T F L O W C O U N T A F T E R A _ _ _ M I N U T E D E L A Y 1 9 9 9 R U N E X T E N D E D 1 P R O G R A M V I E W R E P O R T O T H E R F U N C T I O N S NO S T A R T F L O W C O U N ...

Page 191: ...S A M P L E S P E R B O T T L E 1 m a x 2 0 0 m l S A M P L E S D I S T R I B U T I O N B O T T L E S S A M P L E R U N C O N T I N U O U S LY _ _ B O T T L E S P E R S A M P L E E V E N T 1 m a x S E Q U E N T I A L B O T T L E S S A M P L E S A M P L E S B O T T L E Multi Bottle 1 Bottle No S E L E C T U N I T S F O R F L O W R A T E c f s g p s g p m M g d l p s m 3 s m 3 h m 3 d S E L E C T U ...

Page 192: ...I N U T E D E L A Y 1 9 9 9 S E L E C T D A Y S S U M O T U W E T H F R S A D O N E F I R S T S A M P L E H H M M M O T H F I R S T S A M P L E AT H H M M P R O G R A M M I N G C O M P L E T E R U N T H I S P R O G R A M N O W Y E S N O R U N P R O G R A M V I E W R E P O R T O T H E R F U N C T I O N S NO YES A T L E A S T O N E D A Y M U S T B E S E L E C T E D D O N E If applicable See Section ...

Page 193: ... C O M B I N E D R E S U L T S B O T H N O N E R E P O R T S F O R R A I N D A I LY S U M M A R Y C O M B I N E D R E S U L T S B O T H N O N E D O Y O U W A N T C O M B I N E D R E S U L T S F O R S D I 1 2 D A T A Y E S N O I N C L U D E D E T A I L D A T A F O R F L O W L I N K A N D T E X T L I N K P R O G R A M S Y E S N O R E P O R T S F O R F R T E M P D A I LY S U M M A R Y C O M B I N E D...

Page 194: ... _ R E I N I T I A L I Z E Y E S N O D I D Y O U R E P L A C E T H E I N T E R N A L B A T T E R Y Y E S N O P L E A S E W A I T T E S T I N G R A M R A M P A S S E D T E S T P L E A S E W A I T T E S T I N G R O M R O M PA S S E D T E S T A B C D E F G H I J K L M N O P Q R S T A B C D E F G H I J K L M N O P Q R S T A B C D E F G H I J K L M N O P Q R S T A B C D E F G H I J K L M N O P Q R S T ...

Page 195: ... l 1 0 0 9 9 9 0 G R A B S A M P L E P R E S S W H E N R E A D Y C A L I B R A T E V O L U M E P R E S S W H E N R E A D Y V O L U M E D E L I V E R E D _ _ _ _ _ m l _ _ _ _ _ m l A R E Y O U S U R E Y E S N O S E L E C T D I R E C T I O N P U M P F O R W A R D P U M P R E V E R S E G O T O B O T T L E _ _ 1 m a x N O W A T B O T T L E _ _ P U R G I N G P R E S S T H E R E D S T O P K E Y W H E N...

Page 196: ...G E S Y E S N O S A V E C H A N G E S Y E S N O E X T E N D E D 1 E X T E N D E D 2 E X T E N D E D 3 E X T E N D E D 4 N O N O Y E S P R O G R A M M O D U L E Y E S N O S E L E C T U N I T S F O R L E N G T H f t m N E W M O D U L E S E T U P D O W N L O A D D A T A N O W O R L O S E A L L D A T A D O N E I f a p p l i c a b l e S E L E C T U N I T S F O R F L O W R A T E c f s g p s g p m M g d ...

Page 197: ...9 9 A U T O S U C T I O N H E A D E N T E R H E A D _ R I N S E C Y C L E S 0 3 R E T R Y U P T O _ T I M E S W H E N S A M P L I N G 0 3 O N E P A R T P R O G R A M T W O PA R T P R O G R A M _ _ B O T T L E S A V A I L A B L E A S S I G N B O T T L E S 1 T H R U _ _ T O P A R T A 1 m a x S U C T I O N H E A D O F _ _ f t m i n m a x Continued from Figure A 9 Extended Programming Programming Scre...

Page 198: ... E S A M P L E S A T 1 S T A R T T I M E 2 H H M M 3 H H M M T A K E _ _ _ S A M P L E S 1 9 9 9 R U N C O N T I N U O U S LY Y E S N O R U N C O N T I N U O U S LY Y E S N O D O Y O U W A N T T O R U N C O N T I N U O U S LY Y E S N O _ _ B O T T L E S P E R S A M P L E E V E N T 1 m a x S W I T C H B O T T L E S E V E R Y _ _ _ S A M P L E S 1 m a x S W I T C H B O T T L E S O N N U M B E R O F ...

Page 199: ...I N U T E S L E V E L R A N G E L O W E R _ _ _ f t U P P E R _ _ _ f t 0 0 0 1 3 0 0 0 0 L E V E L S E T P O I N T _ _ _ f t 0 0 0 1 3 0 0 0 0 E N A B L E D W H E N I N S I D E R A N G E O U T S I D E R A N G E E N A B L E D W H E N A B O V E S E T P O I N T B E L O W S E T P O I N T L E V E L C O N D I T I O N S E T P O I N T R A N G E R A T E O F C H A N G E R E S E T R A I N H I S T O R Y Y E ...

Page 200: ... 0 0 0 1 3 0 0 0 0 L E V E L S E T P O I N T _ _ _ f t 0 0 0 1 3 0 0 0 0 E N A B L E D W H E N I N S I D E R A N G E O U T S I D E R A N G E E N A B L E D W H E N A B O V E S E T P O I N T B E L O W S E T P O I N T E N A B L E D W H E N I O E N A B L E I S H I G H I O E N A B L E I S L O W R E S E T R A I N H I S T O R Y Y E S N O R E P E A T F O R S E C O N D T Y P E I O All except I O Rain and A...

Page 201: ...H f t U N I T S S E L E C T E D F L O W R A T E c f s B U B B L E R M O D U L E W E I R 9 0 o V N O T C H F L O W V O L U M E M g a l I f a p p l i c a b l e S E L E C T U N I T S F O R F L O W R A T E c f s g p s g p m M g d l p s m 3 s m 3 h m 3 d F L O W V O L U M E c f g a l M g a l m 3 l i t S E L E C T U N I T S F O R N E W M O D U L E S E T U P D O W N L O A D D A T A N O W O R L O S E A L ...

Page 202: ...E N G T H I S _ _ f t 3 9 9 A U T O S U C T I O N H E A D E N T E R H E A D S U C T I O N H E A D O F I S _ _ f t m i n m a x _ R I N S E C Y C L E S 0 3 R E T R Y U P T O _ T I M E S W H E N S A M P L I N G 0 3 O N E P A R T P R O G R A M T W O P A R T P R O G R A M _ _ B O T T L E S A V A I L A B L E A S S I G N B O T T L E S 1 T H R U _ _ T O P A R T A 1 m a x O N E P A R T P R O G R A M When L...

Page 203: ...T A R T T I M E 2 H H M M 3 H H M M T A K E _ _ _ S A M P L E S 1 9 9 9 _ _ B O T T L E S P E R S A M P L E E V E N T 1 m a x S W I T C H B O T T L E S E V E R Y _ _ _ S A M P L E S 1 m a x S W I T C H B O T T L E S O N N U M B E R O F S A M P L E S T I M E F I R S T S W I T C H T I M E A T _ _ _ _ S W I T C H B O T T L E S E V E R Y _ _ H O U R S _ _ M I N U T E S V O L U M E _ _ _ _ m l S A M P ...

Page 204: ... R D O N E E N A B L E R A I N A N D I O L E V E L F L O W p H T E M P C O N D D O N O N E R E S E T R A I N H I S T O R Y Y E S N O R E P E A T F O R S E C O N D T Y P E I O R A I N A D O N E A l l e x c e p t N O N E D O N E A N D o r O R See Note A See Note A See Note B I f a p p l i c a b l e I f a p p l i c a b l e A l l e x c e p t I O R a i n a n d A D o n e Notes A All valid enable options...

Page 205: ...P U T E V E R Y 1 5 S E C 1 M I N 5 M I N 5 M I N A T T H E B E G I N N I N G O F I N I T I A L P U R G E F O R W A R D P U M P I N G D U R I N G E N T I R E P U M P C Y C L E F O R W A R D P U M P I N G M E A S U R E M E N T I N T E R V A L 5 1 5 3 0 S E C O N D S 1 2 5 M I N U T E S S T O R A G E I N T E R V A L P R O G R A M M I N G S T Y L E N O R M A L Q U I C K V I E W C H A N G E E V E N T ...

Page 206: ...B A C K L I G H T P R O G R A M M I N G S T Y L E N O R M A L Q U I C K V I E W C H A N G E M E A S U R E M E N T I N T E R V A L 5 1 5 3 0 S E C O N D S 1 2 5 M I N U T E S S T O R A G E I N T E R V A L E V E N T M A R K D U R A T I O N 3 S E C O N D S W H I L E P U M P I N G A T T H E B E G I N N I N G O F I N I T I A L P U R G E F O R W A R D P U M P I N G D U R I N G E N T I R E P U M P C Y C ...

Page 207: ...U N E R R O R F L O W T H R U C O N D I T I O N S See Note A R A I N G A U G E 0 0 1 i n c h T I P 0 1 m m T I P Y E S N O S C A N S D I 1 2 S O N D E S A T T U R N O N T I M E D O Y O U H A V E A R A I N G A U G E A T T A C H E D Y E S N O Y E S N O N E W H A R D W A R E S E T U P If applicable P A R T I T I O N M E M O R Y W I L L L A S T _ _ D A Y S P L E A S E W A I T Notes A denotes input sig...

Page 208: ...O R L O S E A L L D A T A D O N E Y E S NO N O S D I 1 2 S O N D E A U T O S D I 1 2 S C A N O N If applicable Notes A denotes input signal denotes output signal R U N F L O W T H R U P U M P _ _ _ S E C O N D S B E F O R E T A K I N G R E A D I N G S 0 1 2 0 S E T I O x N O N E I O E N A B L E F R T E M P P G M R U N N I N G P G M E N A B L E D A E N A B L E D B E N A B L E D P R O G R A M D O N ...

Page 209: ... s p r e s s e d I O x 1 2 I O x 3 S e e N o t e B S E T U P M A N U A L C O N T R O L A N A L O G O U T P U T _ L E V E L V E L O C I T Y F L O W N O N E S E T A N A L O G O U T P U T _ T O _ _ _ _ m A D O N E A N A L O G O U T P U T _ 0 T O 2 0 m A 4 T O 2 0 m A A N A L O G O U T P U T _ 0 m A AT _ _ _ f p s 2 0 m A AT _ _ _ f p s 9 9 9 9 9 9 9 9 9 9 D I A L O U T W H E N I O x B E C O M E S T R...

Page 210: ... __ mA DONE ANALOG OUTPUT _ 0 TO 20 mA 4 TO 20 mA ANALOG OUTPUT _ 0 mA AT _ __ fps 20 mA AT _ __ fps 99999 99999 _ ANALOG OUTPUTS Repeat for each channel A Factory installed option Requires B1 hardware or newer Note I Ox 1 2 I Ox 3 DIALOUT WHEN I Ox BECOMES TRUE YES NO DIALOUT NUMBERS __ MINUTE DELAY BETWEEN CALLS 0 99 DIALOUT EACH TIME I Ox BECOMES TRUE 1 2 3 When DECIMAL POINT key is pressed SEL...

Page 211: ... A T A N O W O R L O S E A L L D A T A D O N E S E L E C T A D D R E S S O F S O N D E T O S E T U P _ _ _ _ D O N E C o m p a n y M o d e l O p t i o n a l I n f o r m a t i o n S O N D E A T A D D R E S S _ H A S _ D A T A V A L U E S If No Sondes Found If Sondes Found Non Isco Ready Sondes Done Not enough partitions to store this sonde s data Repeat for all data types Enough partitions to store...

Page 212: ...s This appendix provides Material Safety Data Sheets for the des iccant used by the BLZZRD Sampler Teledyne ISCO cannot guarantee the accuracy of the data Spe cific questions regarding the use and handling of the products should be directed to the manufacturer listed on the MSDS ...

Page 213: ...ravity H20 1 2 1 Vapor Pressure mm Hg and Temperature N A Melting Point N A Vapor Density Air 1 N A Evaporation Rate __________ 1 N A Solubility in Water Insoluble but will adsorb moisture Water Reactive Not reactive but will adsorb moisture Appearance and Odor Purple crystals no odor Section 3 Fire and Explosion Hazard Data Flash Point and Methods Used N A Auto Ignition Temperature N A Flammabili...

Page 214: ...Ingestion Drink at least 2 glasses of water Section 6 Control and Protective Measures Respiratory Protection Specify Type Use NIOSH approved dust mask or respirator Protective Gloves Light cotton gloves Eye Protection Safety glasses VENTILATION Local Exhaust Mechanical General Special TO BE USED Other Specify Other Protective Clothing and Equipment None Hygienic Work Practices Avoid raising dust A...

Page 215: ...PEL Amorphous Silica 63231 67 4 99 PEL 20 RESPIRABLE TLV 5 LIMIT NONE HAZARD IRRITANT Synthetic amorphous silica is not to be confused with crystalline silica such as quartz cristobalite or tridymite or with diatomaceous earth or other naturally occurring forms of amorphous silica that frequently contain crystalline forms This product is in granular form and packed in bags for use as a desiccant T...

Page 216: ...ing sound from venting safety device or any discoloration of the tank due to fire SECTION V HEALTH HAZARD DATA Health hazards may arise from inhalation ingestion and or contact with the skin and or eyes Ingestion may result in damage to throat and esophagus and or gastrointestinal disorders Inhalation may cause burning to the upper respiratory tract and or temporary or permanent lung damage Prolon...

Page 217: ... personnel need protection against inhalation of dusts or fumes Eye protection is required Vacuuming and or wet methods of cleanup are preferred Place in appropriate containers for disposal keeping airborne particulates at a minimum SECTION VIII SPECIAL PROTECTION INFORMATION Respiratory Protection Provide a NIOSH MSHA jointly approved respirator in the absence of proper environmental control Cont...

Page 218: ...ation contained herein is based upon data considered true and accurate However United Desiccants makes no warranties expressed or implied as to the accuracy or adequacy of the information contained herein or the results to be obtained from the use thereof This information is offered solely for the user s consideration investigation and verification Since the use and conditions of use of this infor...

Page 219: ...BLZZRD Portable Refrigerated Sampler Appendix B Material Safety Data Sheets B 8 ...

Page 220: ...on of safety procedures offers only general guidelines Each situation in which you install a sampler varies You must take into account the individ ual circumstances of each installation Additional safety considerations other than those discussed here may be required Check applicable guidelines codes and regulations of federal state city and county agencies C 1 Hazards There are many hazards connec...

Page 221: ...the atmosphere in the manhole should be retested before anyone enters the manhole When explosive conditions are encountered the ventilating blower should be placed upwind to prevent igniting any gas that is emerging from the opening When a gasoline engine blower is used it must be located so that exhaust fumes cannot enter the manhole If testing equipment is not available the manhole should be ass...

Page 222: ...ncoming traffic Flashing warning signals should be used to alert drivers and pedestrians Orange safety vests should be worn by personnel stationed at the surface when the manhole is located in a vehicular traffic area C 4 2 Removing the Covers Manhole covers should be removed with a properly designed hook Use of a pick ax screwdriver or small pry bar may result in injury A suitable tool can be mad...

Page 223: ...the earlier years he did so he admits with little knowledge of the grave hazards to which he exposed himself It is gratifying that the subject of hazards to people work ing in sewers is receiving much more attention than in past years and good safety procedures are prescribed in various publications on this subject It is essential that people know and use correct procedures It is less important to...

Page 224: ...le reduction in the oxygen concen tration Waste water containing 2 mg per liter of dissolved sulfide and at a pH of 7 0 can produce in a chamber with high turbulence a concentration of 300 PPM H2S in the air This is considered to be a lethal concentration Many people have died from H2S not only in sewers and indus tries but also from swamps and from hot springs In one resort area at least five per...

Page 225: ...armful vapors They too are sensed by smell and explosimeter tests if they get into the public sewer Such occurrences are rare The attempt to instill a sense of urgency about real haz ards is diluted if a man is told to give attention to a long list of things that in fact are irrelevant Be very careful to avoid high H2S concentrations flamma ble atmospheres and hazards of physical injuries Remem be...

Page 226: ...O Chemical asphyxiant Col orless odorless tasteless Flammable Poisonous 0 97 Combines with hemoglobin of blood Unconsciousness in 30 min at 0 2 to 0 25 Fatal in 4 hours at 0 1 Headache in few hours at 0 02 400 50 12 5 74 0 Near top especially if present with illuminating gas Manufactured gas flue gas products of combustion motor exhausts Fires of almost any kind CO ampoules Carbon Tetra Chloride C...

Page 227: ...oor ventilation and absorption or chemical consump tion of oxygen Oxygen deficiency indicator Ozone O3 Irritant and poisonous Strong electrical odor Strong oxidizer Colorless At 1 ppm strong sulfur like odor 1 66 Max naturally occurring level is 0 04 ppm 0 05 ppm causes irritation of eyes and nose 1 to 10 ppm causes headache nausea can cause coma Symptoms sim ilar to radiation damage 0 08 0 04 Nea...

Page 228: ...er to the tables to determine the part number for the item Replacement parts can be purchased on our online store store teledyneisco com or by contacting Teledyne ISCO s Customer Service Department Teledyne ISCO Customer Service Department P O Box 82531 Lincoln NE 68501 USA Phone 800 228 4373 402 464 0231 FAX 402 465 3022 E mail IscoInfo teledyne com ...

Page 229: ...27 1 8 liter glass bottles and caps Set of 8 60 3004 193 Cap with liner for 2 5 Gallon Glass bottle Photo Coming Soon BLZZRD 1 bottle parts 60 9004 051 Tube guide 68 2700 005 2 5 gallon 9 5 liter glass round bottle with PTFE lined cap 60 3004 193 Cap with liner for 2 5 Gallon Glass bottle 68 2740 009 2 5 gallon 9 5 liter polyethylene round bottle with two caps 29 9001 303 lid for 2 5 Gallon Poly b...

Page 230: ...t BLZZRD 6712 Avalanche Pump Shaft Bushing and Seals 60 9004 511 Kit BLZZRD 6712 Avalanche Pump Roller Replacement 60 9004 411 Kit BLZZRD 6712 Avalanche Pump Gear Case Photo Coming Soon Photo Coming Soon Photo Coming Soon Photo Coming Soon Photo Coming Soon 10 9080 300 Draw Latch for Pump Case 60 9004 442 Kit BLZZRD Dust Cap Replacement 60 9004 515 Kit BLZZRD 6712 Avalanche Pump Gears and Plates 6...

Page 231: ... CPU Replacement Kit 34 0503 004 Coin style Lithium Battery 3 Volt 560 mah 60 2960 012 Display keypad Replacement Kit 60 9004 443 Replacement Module Harness Kit BLZZRD 6712 Avalanche Photo Coming Soon Photo Coming Soon Photo Coming Soon 49 0001 300 Humidity indicator card 60 9004 408 Replacement Connector Circuit Board Kit BLZZRD 6712 Avalanche 60 2960 022 Replacement Refrigeration Control Interna...

Page 232: ...lacement Power Supply Kit 60 2960 014 Replacement Temperature Sensor Kit 60 2960 015 Replacement Compressor Drive Module Kit Photo Coming Soon Photo Coming Soon Photo Coming Soon Photo Coming Soon Photo Coming Soon Photo Coming Soon Photo Coming Soon BLZZRD Top Cover and Controller Mounting Parts 60 2974 012 Top Cover Assembly for BLZZRD Avalanche Sampler 60 2960 019 Replacement Controller Mountin...

Page 233: ...BLZZRD Portable Refrigerated Sampler Appendix D Replacement Parts D 6 ...

Page 234: ...are available To order any item contact your sales representative or the fac tory Teledyne ISCO P O Box 82531 Lincoln NE 68501 USA Phone 800 228 4373 402 464 0231 FAX 402 465 3022 E mail IscoInfo teledyne com Note The part name listed on your order acknowledgment and invoice may be different than the item name listed here When examining these documents use the part number for refer ence ...

Page 235: ...caps two discharge tubes and adapter 68 296 025 BLZZRD 4 bottle configuration Includes 4 glass 1 8 liter bottles with PTFE lined caps distributor arm one discharge tube and adapter plate 68 2960 026 BLZZRD 2 5 gal 10 liter glass bottle with PTFE lined caps Includes two discharge tubes and adapter 68 2960 021 BLZZRD 2 5 gal 10 liter lightweight polyethylene bottle with a cap for sampling and a cap ...

Page 236: ... sold separately 68 6700 052 701 pH Temperature Accessories Stainless Steel double junction combination pH probe with built in temperature probe With 25 ft cable 60 9004 126 720 Submerged Probe Flow Module 720 Flow Module submerged probe Probe mounting hardware sold separately 60 9004 030 720 Submerged Probe Accessories Submerged probe level sensor only measures 10 ft level range With 25 ft connec...

Page 237: ... Low Profile Area Velocity Sensor to Teledyne ISCO Standard Mounting Rings 60 3204 029 780 Analog Input Module 780 analog input module Includes 10 ft cable and instruction manual 60 9004 040 Other Parameter Measurement Equipment 674 Rain Gauge Tips every 0 01 inch of rainfall Tipping bucket rain gauge with 50 ft cable and connector for use with a 4100 Series Flow Logger 4200 Series Flow Meter or 6...

Page 238: ...he amount of energy required to cool the samples correlates to ambient air temperature The higher the air temperature the more the sampler must operate the refrigeration system The equation below may be used as a guide for selecting a battery that will meet the requirements of your sampling program Before using the equation you must know the total sample volume that will be collected the number of...

Page 239: ...er to retrieve it and the ambient temperature is 30 C At this ambient temperature Figure F 1 tells us Ac is 21 and Figure F 2 tells us Am is 41 The equation would be Using the equation we find operating the sampler under these conditions would need a fully charged battery with a capacity of at least 103 amp hours As previously mentioned standby current is negligible In the example above the sample...

Page 240: ...ample liquid from 20 C Ac Figure F 2 Amp hours per day required to maintain sample liquid at 3 C Am Amp hours 0 10 20 30 40 32 77 92 107 62 47 Sample Cooling Demand 122 0 10 20 30 40 50 Ambient F Ambient C Amp hours per Day 0 10 20 30 40 50 60 32 77 92 107 62 47 Sample Maintenance Demand 122 0 10 20 30 40 50 Ambient F Ambient C ...

Page 241: ...BLZZRD Portable Refrigerated Sampler Appendix F Battery Selection Guide F 4 ...

Reviews:

Related manuals for Everywhereyoulook BLZZRD