Instantaneous (updated every 5 seconds) and 1-minute values shown above are stored in three 16 bit
Modbus registers. The first 2 registers contain a floating point number for the parameter. The 3
rd
word
contains a status word with the individual bits define as follows (bit 0 is the least significant bit):
bit #
15
-
not used
14
-
In interference test
13
-
Interference test failure
12
-
Full scale exceeded
11
-
not used
10
-
In Calibration
9
-
fatal fault
8
-
non-fatal fault
7
-
not used
6
-
In Maintenance
5
-
not used
4
-
Communication Failure
3
-
not used
2
-
Data valid
1
-
In Span Calibration
0
-
In Zero Calibration
4.1.5
Data: Tools
This screen provides information on firmware version, IP address, and communication status.
Occasionally the configuration on the Deltaflow controller and what is shown on the web interface may