Operator’s Manual
922258-00 Rev A
45
Rejecting the AC line voltage was no real challenge, but rejecting the drain to source signal is a real
measure of system ability. The drain to source voltage rises over 245 Volt when the FET turns OFF.
The maximum rate of rise of this signal is about 15 V/nsec followed by a ring at the bottom of the
waveform. It will be necessary to adequately reject this signal if upper gate signal is the be
measured accurately.
Upper and Lower Gate Drive
To examine the gate drive signal on the upper FET’s Q1, the –INPUT probe will be connected to the
source of Q1 and the +INPUT probe to the gate of Q1. The XC100 probes are set for an attenuation
of ÷100 and the DA1855A for an attenuation of ÷1 and a gain of X1. The EFFECTIVE GAIN indicator
should read an overall gain of ÷100. To make room for other traces, the OFFSET control on the
oscilloscope was set to –5.0 Volt, moving the trace up one division.
The same procedure is repeated for Q2 gate drive where the OFFSET is set to +15 V to move the
trace down by 3 divisions. By setting mVOLT/DIV
to a more sensitive setting, small details of these
signals can be examined.
Avoiding Measurement Errors
The math capabilities in modern digital oscilloscopes can save time and effort. Both scalar
measurements and waveform math provide direct answers for measurements that used to require
considerable computation and analysis. Common causes of erroneous results are:
•
Errors in conditioning the input signal, such as clipping or bandwidth limiting.
•
Limitation in the acquisition process, such as sample rate, resolution and record length.
•
Limitation in the computational algorithms.
The most common source of error in power measurements results from the time delay (skew)
between the voltage and current waveforms. The propagation delay through the current probe and
the voltage probe plus differential amplifier are almost never equal. To eliminate resulting error in
power waveforms, it is necessary to deskew the input signals. Some oscilloscopes have a deskew
function that can be used to shift the time reference of one of the waveforms relative to the other.
Another error to be concerned with is the phase shift in the probes or instrument. As the rise time
of the input signal approaches the rise time of the current probe or amplifier, the phase shift will
create an amplitude error in the power waveform.
Summary of Contents for DA1855A
Page 1: ...Operator s Manual DA1855A Differential Amplifier ...
Page 2: ......
Page 3: ...DA1855A Differential Amplifier Operator s Manual February 2013 ...
Page 93: ......
Page 94: ......