CombiFlash® Rf
+
User Manual
2-24
Use ELSD by Default
- causes all runs to use the ELSD
detector unless disabled in the run requirement screen
[S
2.13.8 Prep HPLC Tab
The
DEFAULT
COLUMN
allows selection of the most commonly
used Preparative HPLC column. This will be the method
loaded when the user first selects use of the Preparative HPLC
function during normal operation.
Many times this will be the
column mounted onto the system. The drop down list to select
the
DEFAULT
COLUMN
is based on the columns already defined
by the user. To define a column, refer to the area below the
DEFAULT
COLUMN
field.
To define a column:
1. Press N
EW
. This will open a dialog box allowing the new
column to be named. This name will be used to refer to the
column in normal usage. This name will also be printed on
the report.
2. Select the
MEDIA
SIZE
. This data will be printed on the
chromatograms for future reference.
3. Enter a
COLUMN
DIAMETER
. This information will be used
by the system to recommend an optimal flow rate to
achieve highest resolution for the column being defined.
4. Enter a
COLUMN
LENGTH
. This is used to recommend a
separation length. This length is based on a gradient
segment approximately 12 column volumes in duration.
5. Select
DEFINE
METHOD
. This allows a review of the method
defined by the system. This is an opportunity to edit the
recommended gradient shape and duration to better match
the expected chemistry. A higher flow rate can be selected
to balance resolution and throughput.
6. Selecting S
AVE
. This will save the column and allow its use
on the system.
7. If you are done defining columns, press the
OK
button.
If you have several columns that will be used on the system,
these can be defined at this time or at any later time.
• S
AVE
A
S
allows you create a new column definition
based on an existing column. This is done by selecting
the existing column in the C
OLUMN
N
AME
drop down
Summary of Contents for CombiFlash EZ Prep
Page 2: ...CombiFlash Rf User Manual ii ...
Page 8: ...CombiFlash Rf User Manual viii ...
Page 52: ...CombiFlash Rf User Manual 2 16 Figure 2 7 Loading test tubes ...
Page 85: ...CombiFlash Rf User Manual 3 20 ...
Page 103: ...CombiFlash Rf User Manual 4 18 ...
Page 185: ...CombiFlash Rf User Manual 6 64 ...
Page 199: ...CombiFlash Rf User Manual 7 14 ...
Page 223: ...CombiFlash Rf User Manual 8 24 ...
Page 225: ......
Page 227: ......
Page 229: ......
Page 231: ......
Page 233: ......
Page 235: ......
Page 237: ......