Automation
Section 2 The AutoInjector Module
2-9
Figure 2-14 Configuring the injection loop volume
For best results, the loop should generally be 50% filled for
maximum compound recovery. If the parameters entered for the
individual injection exceed a 50% loop fill, a warning message
will appear recommending a change in the parameters. This
warning can be ignored at the discretion of the user.
2.2.2 Operation
To use the AutoInjector Module, follow the steps below:
1. Place the sample probe into your sample container.
2. Press P
LAY
to start a separation as you would for an indi-
vidual separation.
3. If you want to automatically perform multiple injections
(or perform a single injection using the AutoInjector Mod-
ule), select the “Multiple Injections” drop down as shown in
Figure 2-15. This is the default selection if an AutoInjector
Module is present.
4. If you want to manually inject your sample without using
the AutoInjector Module, select “Load Now” or “Load After
Equilibrium”. When prompted to load the sample, remove
the sample probe line from the inject valve and inject
directly into the valve.
Summary of Contents for AutoInjector Module
Page 2: ......
Page 6: ...Automation Modules Installation and Operation Guide Safety Warnings 6 ...
Page 8: ...Automation Modules Table of Contents 8 ...
Page 10: ...Automation Modules Section 1 Introduction 1 2 ...
Page 46: ...Automation Section 3 The AutoSampler Module 3 24 ...
Page 64: ...Automation Section 4 The Column Selector Module 4 18 ...
Page 68: ...Automation Appendix A Specifications A 4 ...
Page 69: ...Compliance Statements C E D e c l a r a t i o n o f C o n f o r m i t y P e n d i n g ...
Page 70: ......
Page 72: ......
Page 74: ......