background image

Information included herein is controlled by the Export Administration Regulations (EAR) and requires an export license, license exception or 

other approval from the appropriate U.S. Government agency before being exported from the United States or provided to any foreign person. 

Diversion contrary to U.S. law is prohibited.

ACCQPrep Pump Assembly 

Replacement and HP150 to  

HP125 Pump Assembly      

Conversion Instructions

Instruction Sheet #69-5233-940

Revision A, Jan 2, 2020

Overview

The ACCQPrep replacement pump assembly contains 
a pump sled assembled for a HP150 instrument, parts 
to convert the pump assembly to be used on a HP125 
instrument, and parts to convert the old Valco pump 
bulkhead brackets and fittings to Swagelok on older 
HP125 instruments. If the replacement pump assembly 
is being installed on an HP150 configured ACCQPrep, 
then no conversion work needs to be completed before 
installing the replacement pump assembly.

Table 1: Parts List

Item

Quantity

Replacement Pump Assembly

1

Swagelok Bulkhead Assembly

1

The table below lists additional items that are required 
for the pump assembly conversion:

Table 2: Required Tools

Safety

The use of Personal Protective Equipment (PPE) is highly 
encouraged due to potential contact with residual solvents.

Time Required:

Typical time required to perform the pump greasing is 
approximately 1.5 hour.

  

 Note

If the Swagelock Bulkhead Assembly (60-5237-141) will  
be installed with the replacement pump, then use these  
instructions to complete the work and ignore the 
installation instructions (69-5233-941) that come with 
the Swagelok Bulkhead Assembly.

Removing Solvent & Powering Down  

the Instrument

1.  Remove all solvent supply lines from the solvent 

containers.

2.  Remove the lower front cover of the instrument to 

expose the front of the current pump assembly

3. Select 

Tools > Manual Control

 from the upper tabs  

of the main screen of the instrument.

4.  Set the 

Flow Path

 to “Prep HPLC column”. Set the 

Flow Rate, ml/min

 to 50.

5.  If the instrument does not include a Solvent Select Valve, 

then proceed to Step 6, else Set 

Solvent A

 to one of the 

plumbed solvent lines and 

Solvent B

 to another of the 

plumbed solvent lines (ex: A1 - Methanol, B1-Water, etc.).

6.  Press the 

Pump Solvent A

 button. Allow the pumps to 

run until there is no longer solvent flowing to the bottom 
of the A pumps. Press the 

Stop

 button. Press the 

Pump 

Solvent B

 button and repeat this process for the B pumps.

7.  Shutdown the instrument by toggling the red power 

switch on the front of the instrument.

8.  Wait 30 seconds after the touchscreen turns off and 

then disconnect power to the instrument either by 
disconnecting the power cable at the back of the 
instrument or at the wall outlet.

Removal of the Current Pump Assembly

1.  With your fingers, disconnect the input tube at 

Y-connector below each set of pumps (Figure 1). If the 
fittings are too tight to be disconnected by hand, then 
gently use a pliers to loosen the fitting enough to finish 
disconnecting it by hand.

 

Figure 1:  Y-Fitting inlet tubing

Item
#2 Phillips Screwdriver
1/8” Small Flathead  
Screwdriver
5/16” Open End Wrench
1/4” Open End Wrench

Side Cutter
5/16” Nut Driver or  
Rachet (optional)
Pliers
1/4” Rachet (optional)

Reviews: