5.8 Software Updates
Update Isco Instrument Software installs software in Isco instru-
ments and modules. This application will transfer a binary file
from your PC to the instrument's flash memory.
Updated binary files are available from Teledyne Isco when
enhancements have been made to the instrument software.
Contact the factory for help with obtaining these files.
CAUTION
Updating the instrument software or resetting the instru-
ment may completely erase readings and reports stored in
the instrument's memory. Collect the data from the instru-
ment before proceeding.
5.8.1 Updating the Flow
Meter Software
The flow meter should be powered and turned on.
Connect your computer to the flow meter using the Computer
Connect Cable (Interrogator Cable) and start the program
Update Isco Instrument Software.
Figure 5-10 Open the software update program
Click the Help button and follow the steps listed under Update
Instrument Software. Do not disturb the connection or stop the
update while the file transfer is taking place. When the transfer
is complete, the program displays an “Operation successful”
message. Click OK and close the program.
Your flow meter’s software is now updated and ready for oper-
ation.
Summary of Contents for 4250
Page 2: ......
Page 4: ......
Page 12: ...4250 Flow Meter Table of Contents x ...
Page 28: ...4250 Flow Meter Section 1 Introduction 1 16 ...
Page 90: ...4250 Flow Meter Section 2 Programming 2 62 ...
Page 140: ...4250 Flow Meter Section 4 Options and Accessories 4 30 ...
Page 160: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 2 ...
Page 161: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 3 ...
Page 166: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 8 ...
Page 184: ...4250 Flow Meter Appendix C General Safety Procedures C 10 ...
Page 191: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 7 ...
Page 192: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 8 ...
Page 196: ......
Page 198: ......
Page 200: ......
Page 202: ...Warranty ...
Page 204: ......