Teledyne API Model 401 O
3
Photometric Calibrator Instruction Manual, 01124, Rev. J1
2.
Confirm that all jumpers on the V/F board are set properly, as follows:
Table 9-3: V/F Board Jumpers - Factory Settings
Factory Set Jumpers
Jumper
Setting
B1 1
B2 1-2
B12 3-4
(0-5V)
B14 2-3
(Enabled)
B15
Set to match power line frequency
JP1 1-2
JP2 1-2
Table 9-4: V/F Board Dip Switches - Ranges For Analog Output
User Set DIP Switches
Switch
100 mV
Full Scale
1 V
Full Scale
5 V
Full Scale
10 V
Full Scale
S1(DAC 0)= “REC”
1,6 = ON
1,5 = ON
1,4 = ON
1,3 = ON
S2(DAC 1)= “DAS”
1,6 = ON
1,5 = ON
1,4 = ON
1,3 = ON
S3(DAC 2)=”TEST”
1,6,7 = ON
1,5,7 = ON
1,4,7 = ON
1,3,7 = ON
S3(DAC 2)=”O3 GEN”
N/A
N/A
1,4,7 = ON
N/A
3.
If Voltages and Jumper settings are correct, the V/F card is faulty and should be replaced.
9-12