Teledyne API Model 401 O
3
Photometric Calibrator Instruction Manual, 01124, Rev. J1
Table 7-2: Warning Message Clear Commands
Command
Warning Message Cleared
"W WSYSRES<CRLF>"
SYSTEM RESET
"W WRAMINIT<CRLF>"
RAM INITIALIZED
"W WALMPINT<CRLF>"
ANA LAMP WARNING
"W WO3LMPHLT<CRLF>"
O3 GEN LAMP SHUTDOWN
"W WO3REFDET<CRLF>"
O3 GEN DETECTOR WARNING
"W WO3FLOW<CRLF>"
O3 GEN FLOW WARNING
"W WSMPFLOW<CRLF>"
SAMPLE FLOW WARNING
"W WSMPPRES<CRLF>"
SAMPLE PRESSURE WARNING
"W WSMPTEMP<CRLF>"
SAMPLE TEMP WARNING
"W WBOXTEMP<CRLF>"
BOX TEMP WARNING
"W WO3LMPTMP<CRLF>"
O3 GEN LAMP TEMP WARNING
"W WALMPTMP<CRLF>"
ANA LAMP TEMP WARNING
"W WVFINS<CRLF>"
V/F NOT INSTALLED
7.3 Status/Control
This subset of messages is concerned with reporting the status of the calibrator and controlling
the calibrator remotely. Whenever the calibrator does a calibration it issues a report to the RS-
232 output. The table on the following page summarizes the status reports.
Table 7-3: Status Reports
Report
"C DDD:HH:MM IIII START ZERO CALIBRATION"
"C DDD:HH:MM IIII FINISH ZERO CALIBRATION"
"C DDD:HH:MM IIII START SPAN CALIBRATION"
"C DDD:HH:MM IIII FINISH SPAN CALIBRATION"
"C DDD:HH:MM IIII START MULTI-POINT CALIBRATION"
"C DDD:HH:MM IIII FINISH MULTI-POINT CALIBRATION"
"C DDD:HH:MM IIII START CALIBRATION HOLD"
"C DDD:HH:MM IIII FINISH CALIBRATION HOLD"
7-4