3.5 Summary Of Front Panel Calibration Controls
The calibration controls are summarized below in terms of the button sequences used to access
them.
Table 3-1: Calibration Controls
Button Sequence
Function
Default
Limits
STBY
Select stand by mode
-
-
ZERO-ENTR
Begin zero calibration
-
-
O3GN-ENTR
Begin span calibration
-
-
ASEQ-ENTR
Begin multi-point calibration
-
-
SETUP-WAIT
Zero/span/duration
15 min
1-20 min
SETUP-EVNT-TIME
Auto check time-of-day
23:30
00:00-23:59
SETUP-EVNT-ASEQ
En/disable auto multi-point check
OFF
OFF/ZS/Z2S/Z5S
SETUP-EVNT-SHFT
Auto check time shift
0 min
SETUP-MISC-O3-
GEN-MODE
Select lamp control mode
Bench
±
60 min
CNST/REF/BNCH
3.6 Photometer Calibration
Model 401 is capable of accurately reproducing O
3
span gas. However, it should be
quantitatively verified by relating to a master standard such as certified Model 400 O
3
photometric analyzer. The calibration of Model 401 photometer can be verified by periodic
(recommended every 3 months) intercomparison between Model 401 photometer and a master
standard.
Photometer Zero Calibration
1.
Turn the power on for both Model 401 and Model 400 (as master standard) and allow them
to stabilize for minimum of one hour.
2.
Start to generate the zero gas by pressing "zero" button from the main menu and enter the
password (512). In case of using external dry zero air, verify proper IZS flow and pressure of
external air source. (See Figure 1-5 for pneumatic connections.)
3-3