3700 Portable Samplers
Section 3 Programming Guidelines
3-37
each pumping cycle. Then, check the suction-line entries in the
configure sequence to see that they are accurate. The calibration
procedure is intended to be for “fine tuning” only.
After you enter the sample volume actually delivered, as shown
in Step 6, all subsequent sample volumes delivered will be
adjusted to correct for the difference between the expected
sample volume and the actual volume entered. Clear the
adjustment by changing a suction line entry in the Suction Line
configure option or by reinitialization of the sampler.
Because the sample volume can be calibrated to ± 10 ml, use a
graduated cylinder to facilitate measurement. Teledyne Isco
offers a graduated cylinder and spare bottles as accessories from
the factory. Refer to the Appendix B Accessories List for details.
The calibration pump cycle will include rinse cycles and retries, if
the sampler is configured to perform them. This ensures that the
calibration procedure includes the pump cycle used while
running the sampling program.
Example: Calibration
1.
Press
ENTER
/
PROGRAM
to access the interactive state.
. . . STANDBY . . .
9:34:50 19-JUN-04
2.
Access the program sequence by selecting “PROGRAM.”
[
PROGRAM
, CONFIGURE]
SAMPLER
3.
Step through the program until the “CALIBRATE SAMPLER?” input display appears (Step
4
).
[
TIME
, FLOW]
PACED SAMPLING
Other program sequence displays.
↓
↓
4.
Select “YES.”
CALIBRATE SAMPLER?
[
YES
, NO]
5.
Before pressing the
MANUAL
SAMPLE
key, make sure a collection container is underneath the distributor.
PRESS MANUAL SAMPLE
KEY WHEN READY . . .
The sampler will deliver the programmed sample volume.
. . . MANUAL SAMPLE . . .
PUMPING 200
ml
6.
Measure the actual volume delivered and enter that value here.
200
ml VOLUME
DELIVERED
7.
Repeat the procedure if desired by selecting “YES.” When you are satisfied with the calibration, select “NO.”
Under normal conditions, you should not need to repeat the procedure.
CALIBRATE SAMPLER?
[YES,
NO
]
Summary of Contents for 3700
Page 2: ......
Page 4: ......
Page 33: ...3700 Portable Samplers Section 2 Setup Procedures 2 9 Figure 2 7 Locked Sampler ...
Page 34: ...3700 Portable Samplers Section 2 Setup Procedures 2 10 ...
Page 144: ...3700 Portable Samplers Section 3 Programming Guidelines 3 110 ...
Page 152: ...3700 Portable Samplers Section 4 Options and Interfacing Equipment 4 8 ...
Page 164: ...3700 Portable Samplers Section 5 Routine Maintenance 5 12 ...
Page 180: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 2 ...
Page 181: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 3 ...
Page 182: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 4 ...
Page 183: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 5 ...
Page 184: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 6 60 3703 262 REV J 3 2 ...
Page 186: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 8 ...
Page 188: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 10 ...
Page 189: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 11 ...
Page 190: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 12 ...
Page 191: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 13 ...
Page 192: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 14 ...
Page 193: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 15 ...
Page 194: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 16 ...
Page 195: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 17 ...
Page 196: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 18 ...
Page 200: ...3700 Portable Samplers Appendix B Accessories B 4 ...
Page 210: ...3700 Portable Samplers Appendix C General Safety Procedures C 10 ...
Page 224: ...3700 Portable Samplers Appendix E Calculating Flow Increment Between Samples E 4 ...
Page 234: ...3700 Portable Samplers Appendix G Material Safety Data Sheets G 8 ...
Page 238: ......
Page 240: ......
Page 242: ......