Frequently Asked Questions & Glossary
Model 200E Instruction Manual
Q:
How often do I need to change the particulate filter?
A:
Once per week. Table 9-1 contains a maintenance schedule listing the most important,
regular maintenance tasks. Highly polluted sample air may require more frequent changes.
Q:
How long does the sample pump last?
A:
The sample pump should last one to two years and the pump head should be replaced
when necessary. Use the
RCEL
pressure indicator on the front panel to see if the pump
needs replacement. If this value goes above 10 in-Hg-A, on average, the pump head needs
to be rebuild.
Q:
Do I need a strip chart recorder or external data logger?
A:
No, the M200E is equipped with a very powerful internal data acquisition system (iDAS).
Section 6.9.10 describes the setup and operation in detail.
Q:
Why does my RS-232 serial connection not work?
A:
There are many possible reasons: 1) the wrong cable, please use the provided or a
generic “straight-through” cable (do not use a “null-modem” type cable), 2) The DCE/DTE
switch on the back of the analyzer is not set properly; make sure that both green and red
lights are on, 3) the baud rate of the analyzer’s COM port does not match that of the serial
port of your computer/data logger. See Section 11.5.11 more trouble-shooting information.
4.2. Glossary
Acronym – A short form or abbreviation for a longer term. Often artificially made up of the
first letters of the phrase’s words.
APICOM – Name of a remote control program offered by Teledyne-API to its customers
ASSY - Acronym for
Assembly.
Chemical formulas used in this document:
•
NO
X
– nitrogen oxides, here defined as the sum of NO and NO
2
•
NO – nitric oxide
•
NO
2
– nitrogen dioxide
•
NO
Y
– nitrogen oxides, often called
odd nitrogen
, the sum of NO, NO
2
(NO
X
) plus
other compounds such as HNO
3
. Definitions vary widely and may include nitrate
(NO
3
-
), PAN, N
2
O and other compounds.
•
NH
3
– ammonia
•
H
2
O – water vapor
•
CO
2
– carbon dioxide
•
SO
2
– sulfur dioxide
•
HNO
3
– nitric acid
DAS - Acronym for
Data Acquisition System
, the old acronym of iDAS
DIAG - Acronym for
diagnostics
, the diagnostic menu or settings of the analyzer
DOC – Acronym for
Disk On Chip
, the analyzer’s central storage area for analyzer operating
system, firmware, and data. This is a solid state device (
!
IC) without mechanical, moving
30
044100102 Rev A
Summary of Contents for 200E
Page 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Page 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Page 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Page 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Page 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Page 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Page 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Page 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...