Model 200E Instruction Manual
APPENDIX C
APPENDIX C: Repair Questionnaire - M200E
Company: _________________ Contact Name: _________ ________________________
Phone Number: _____________ Fax Number: ___________ Email: __________________
Site Address: _______________ _____________________ ________________________
Can we connect to the instrument? If so, provide IP address or modem: ________________
Model 200E Serial Number: ____ _____________________ Firmware revision: _________
The serial number can be found on the back of the instrument, the firmware revision is displayed in the upper left corner of the
display when pressing SETUP on the front panel (Example:
C.3
).
1. List all front panel error/warning messages: __________ ________________________
__________________________ _____________________ ________________________
2. Please complete the following table: (Depending on options installed, not all test
parameters shown below may be available in your instrument)
Parameter
Recorded Value
Acceptable Value
Parameter
Recorded
Value
Acceptable Value
RANGE
ppb/ppm
50 ppb - 20 ppm
PMT TEMP
°
C
7
±
2
NO
X
STB
ppb/ppm
≤
1 ppb with zero air
IZS TEMP
°
C
50
±
1
Sample FL
cm³/min
500
±
50
MOLY TEMP
°
C
315
±
5
Ozone FL
cm³/min
80
±
15
RCEL pressure
in-Hg-A
<10
PMT signal with
zero air
mV
-20 to 150
SAMP pressure
in-Hg-A
Ambient - ~1
PMT signal at
span gas conc
mV
ppb/ppm
0-5000
0-20000 ppb
NO SLOPE
1.0 ± 0.3
NORM PMT at
span gas conc
mV
ppb/ppm
0-5000
0-20000 ppb
NO OFFS
mV
-50 to 150
AZERO
mV
-20 to 150
NO
X
SLOPE
1.0 ± 0.3
HVPS
V
400-900
NOX OFF
mV
-50 to 150
RCELL TEMP
°
C
50
±
1
ETEST PMT
mV
2000
±
1000
BOX TEMP
°
C
A ~5
OTEST
PMT mV
2000
±
1000
3. Has the analyzer been checked for leaks? Yes No ; for proper flows? Yes No
4. What are the failure symptoms? ____________________ ________________________
____________________________ ________________________ ___ Continue on back if necessary
5. Which tests have you done trying to solve the problem? _ ________________________
__________________________ _____________________ __
Continue on back if necessary
6. If possible, fax a portion of a strip chart or email a data file to customer service: 6565
Nancy Ridge Drive, San Diego, CA 92121. Phone: +1 858 657 9800 or 1-800 324 5190.
Fax: +1 858 657 9816. Email: [email protected].
You can access and submit an online version of this form at
http://www.teledyne-api.com/forms/p-fm200E.asp
(04503) 044100102 Rev A
257
Summary of Contents for 200E
Page 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Page 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Page 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Page 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Page 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Page 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Page 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Page 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...