Model 200E Instruction Manual
Theory of Operation
current temperature of the NO
2
converter. It is also stored in the iDAS and reported
as test function
MOLY TEMP
.
•
The sample gas pressure is measured upstream of the reaction cell, stored in the
iDAS and reported as
SAMPLE
. The vacuum gas pressure is measured downstream
of the reaction cell and is stored in the iDAS and reported as
RCEL
.
For more
information on these sensor’s functions see Section 10.2.10.
•
The O
3
gas flow is measured upstream of the ozone generator, stored in the iDAS
and reported as test function
OZONE FL
. For more information on this sensor’s
function see Section 10.2.10.
10.3.5.3. Thermistor Interface
This circuit provides excitation, termination and signal selection for several negative-
coefficient, thermistor temperature sensors located inside the analyzer. They are:
•
The reaction cell temperature sensor is a thermistor embedded in the reaction cell
manifold. This temperature is used by the CPU to control the reaction cell heating
circuit and as a parameter in the temperature/pressure compensation algorithm.
This measurement is stored in the analyzer’s iDAS and reported as test function
RCEL TEMP
.
•
BOX TEMPERATURE SENSOR: A thermistor is attached to the motherboard. It
measures the analyzer’s inside temperature. This information is stored by the CPU
and can be viewed by the user for troubleshooting purposes through the front panel
display. It is also used as part of the NO, NO
X
and NO
2
calculations when the
instrument’s Temperature/Pressure Compensation feature is enabled. This
measurement is stored in the analyzer. Memory as the test function
BOX TEMP
and
is viewable
as a test function (Section 6.2.1) through the analyzer’s front panel.
•
The IZS option’s thermistor is attached to the IZS oven and reports the current
temperature of the NO
2
permeation source to the CPU as part of a control loop that
keeps the tube at a constant temperature. It is stored and reported as test function
IZS TEMP
.
•
The thermistor inside the PMT cold block as well as the thermistor located on the
preamplifier board are both converted to analog signals on the preamplifier board
before being sent to the motherboard’s A/D converter. They are not connected to the
motherboard’s thermistor circuitry. The cold block temperature is stored and
reported as test function
PMT TEMP
, whereas the preamplifier board temperature is
neither stored nor reported. It is used only as a reference internal to the
preamplifier.
10.3.5.4. Analog Outputs
The analyzer comes equipped with four analog outputs. On the instrument’s rear panel
ANALOG
connector, they are labeled
A1
,
A2
,
A3
and
A4
.
Outputs labeled
A1
,
A2
and
A3
carry the concentration signals of NO
X
, NO and NO
2
,
respectively. A variety of scaling measurement and electronic factors apply to these signals.
See Section 6.6 for information on setting the reporting range type and measurement range
scaling factors for these output channels. See Sections 6.7 for instructions on setting the
electronic scaling factors for these channels.
044100102 Rev A
179
Summary of Contents for 200E
Page 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Page 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Page 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Page 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Page 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Page 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Page 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Page 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...