Model 200E Instruction Manual
Theory of Operation
ence. For those applications, we recommend to use other analyzer models. Please consult
sales or our website.
10.1.5.3. Light Leaks
The M200E sensitivity curve includes a small portion of the visible light spectrum
(Figure 10-3), hence, it is important to make sure than the reaction cell is completely
sealed with respect to light. To ensure this, all pneumatic tubing leading into the reaction
cell is either opaque (vacuum exit tubing) in order to prevent light from entering the cell or
light penetration is prevented by stainless steel filters and orifices (gas entries).
10.2. Pneumatic Operation
CAUTION
It is important that the sample airflow system is leak-tight and not pressurized
over ambient pressure. Regular leak checks should be performed on the analyzer
as described in the maintenance schedule, Table 9-1. Procedures for correctly
performing leak checks can be found in Section 11.5.
10.2.1. Pump and Exhaust Manifold
NOTE
Relative Pressure versus absolute pressure. In this manual vacuum readings are
given in inches of mercury absolute pressure (in-Hg-A), i.e. indicate an absolute
pressure referenced against zero (a perfect vacuum).
The gas flow for the M200E is created by an external pump (Figure 10-5) that is pneumati-
cally connected through a 6.4 mm / 0.25” tube to the analyzer’s exhaust port located on
the rear panel (Figure 3-2). This pump creates a vacuum of approximately 5 in-Hg-A at one
standard liter/minute, which is provided to various pneumatic components by a vacuum
manifold located just in front of the rear panel (Figure 3-2 and Figure 10-9). Gas flow is
created by keeping the analyzer’s sample gas inlet near ambient pressure, usually by
means of a small vent installed in the sample line at the inlet, in effect pulling the gas
through the instrument’s pneumatic systems.
There are several advantages to this external pump / pull-through configuration.
•
By using an external pump, it is possible to remove a significant source of acoustic
noise and vibration from the immediate vicinity of the sensor. The PMT can act as a
“microphone”, amplifying noise and vibration within the chassis. This is one of the
main reasons, why the M200E has an external pump.
•
Pumping heats and compresses the sample air, complicating the measurement
process if the pump is upstream.
•
Most importantly, however, certain physical parts of the pump itself are made of
materials that might chemically react with the sample gas. Placing the pump
downstream of the reaction cell avoids these problems.
044100102 Rev A
161
Summary of Contents for 200E
Page 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Page 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Page 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Page 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Page 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Page 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Page 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Page 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...