Teledyne API Model 200A NO
X
Analyzer Instruction Manual, 02246, Rev. F1
Table 5-8: Status Output Pin Assignments
Output #
Pin #
Definition
Condition
1
1,2
ZERO CAL
CLOSED IN ZERO CAL
2
3,4
SPAN CAL
CLOSED IN SPAN CAL
3
5,6
FLOW ALARM
CLOSED IF FLOW WARNING
4
7,8
TEMP ALARM
CLOSED IF ANY TEMP WARNING
5
9,10
DIAG MODE
CLOSED IN DIAG MODE
6
11,12
POWER OK
CLOSED IF SYSTEM POWER OK
7 13,14
SPARE
8 15,16
SPARE
9 17,18
SPARE
10
19,20
AUTORANGE - HI
CLOSED IF IN HIGH RANGE
11
21,22
SYSTEM OK
CLOSED IF NO FAULTS PRESENT
12
23,24
RX CELL PRESS
CLOSED IF ABS PRES > 15" HG
The Status Board schematic can be found in the Appendix.
5.5 RS-232 Interface
The RS-232 communications protocol allows the instrument to be connected to a wide variety of
computer based equipment. The interface provides two basic functions in the M200A.
1.
First is a comprehensive command interface for operating and diagnosing the analyzer in
interactive fashion. This mode is for use by a human operator issuing commands to the
instrument.
2.
Second is the equivalent command interface described above, but with the interface set
up to be operated by a computer program rather than interactive commands.
3.
The interface can also provide an audit trail of analyzer events. In this function the port
sends out messages about instrument events like calibration or warning messages. If these
messages are captured on a printer or remote computer, they provide a continuous audit trail
of the analyzers operation and status.
5-23
Summary of Contents for 200A
Page 10: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK x ...
Page 12: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 1 2 ...
Page 26: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 2 14 ...
Page 40: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 4 10 ...
Page 126: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 7 38 ...
Page 210: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 9 68 ...
Page 218: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 10 8 ...