Teledyne API Model 200A NO
X
Analyzer Instruction Manual, 02246, Rev. F1
3.2 EPA Equivalency Designation
Advanced Pollution Instrumentation, Inc., Model 200A Nitrogen Oxides Analyzer is designated
as Reference Method Number RFNA-1194-099 as defined in 40 CFR Part 53, when operated
under the following conditions:
1.
Range: Any range from 50 parts per billion (ppb) to 1 ppm.
2.
Ambient temperature range of 5 to 40
o
C.
3.
Line voltage range of 105-125 VAC, 60Hz; 220-240 VAC, 50Hz.
4.
With 5-micron TFE filter element installed in the internal filter assembly.
5.
Sample flow of 500 ± 50 cc/min.
6.
Vacuum pump (internal or external) capable of 10"Hg Abs pressure @ 2 slpm or better.
7.
Software settings:
A.
Dynamic span
OFF
B.
Dynamic zero
OFF
C.
Cal-on-NO2 OFF
D.
Dilution factor
OFF
E.
AutoCal
ON or OFF
F.
Independent range
ON or OFF
G.
Autorange
ON or OFF
H.
Temp/Pres compensation ON
I.
Converter Eff.
Acceptable values of 0.96 to 1.02
Under the designation, the Analyzer may be operated with or without the following options:
1.
Rack mount with slides.
2.
Rack mount without slides, ears only.
3.
Rack mount for external pump w/o tray.
4.
Stainless steel zero/span valves.
5.
Internal zero/span.
6.
NO
2
permeation tube - uncertified 0.4ppm @ 0.7 lpm.
7.
NO
2
permeation tube - certified 0.4ppm @ 0.7 lpm.
8.
NO
2
permeation tube - certified 0.8ppm @ 0.7 lpm.
9.
NO
2
permeation tube - uncertified 0.8ppm @ 0.7 lpm.
10.
4-20mA, isolated outputs.
11.
Status outputs.
12.
RS-232 output.
13.
Internal pump or external pump.
14.
1 micron sample filter.
3-2
Summary of Contents for 200A
Page 10: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK x ...
Page 12: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 1 2 ...
Page 26: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 2 14 ...
Page 40: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 4 10 ...
Page 126: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 7 38 ...
Page 210: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 9 68 ...