Teledy
ne API Model 200A NO
X
Analyzer Instruction Manual, 02246, Rev. F1
8-1
8.1 Maintenance Schedule
8
MAINTENANCE
Item
Maintenance Interval
Reference Section
TEST Functions
6 - 12 month intervals
Zero/Span Calibration
Annually or after repairs
Section 7
Zero/Span Checks
Daily
Particulate Filter
Weekly as needed
Ozone Scrubber on Pump Pack
Replace charcoal every 3 months
Section 4.2.7, Figure 4-2
Filter for Ozone Scrubber
Replace every 12 months
Filter for PermaPure Drier
Replace every 12 months
Zero Air Canister
Refill every 3 months
Section 8.5
Reaction Cell Window
Clean annually or as necessary
Section 9.3.8, Figure 8-3
Ozone Flow
Check every year
Figure 9-9, Section 9.3.7
Sample Flow
Check every year
Figure 9-9, Section 9.3.7
Moly Converter
Check efficiency every 6 months
Figure 8-2, Section 8.6
Pneumatic Lines
Examine every 12 months, clean
if necessary
Figure 8-4, Figure 8-5,
Figure 8-6
Factory Calibration
Calibrate each year or after
repairs
Section 9.1.6
Leak Check
Check every year
Section 8.9
O-rings
Replace if seal is broken
Table 8-1: Preventative Maintenance Schedule
NOTE
The operations outlined in this chapter are to be
performed by qualified maintenance personnel only.
Summary of Contents for 200A
Page 10: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK x ...
Page 12: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 1 2 ...
Page 26: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 2 14 ...
Page 40: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 4 10 ...
Page 126: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 7 38 ...
Page 210: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 9 68 ...
Page 218: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 10 8 ...